Cardiac Functions
Fetal / Maternal ECG (Option)
2
2-41
UNIF ACC (uniform acceleration).
5.
Select
RUN
. On the LCD screen, RUNNING flashes while the simulation is active.
6.
A manually generated IUP ends after one contraction. To repeat the simulation, select
RUN
again.
7.
Periodically generated IUPs repeat at the selected period. To terminate a periodic-
IUP simulation, select
END
.
Alternatively, in the Numeric-Control Mode, press the number keys for an IUP-wave
simulation using one of the codes in Table 2-33, and select
RUN
. (The selected IUP is
generated according to the specific period that was last selected, whether manual, or once
every 2, 3, or 5 minutes.)
Table 2-33. Numeric Codes for IUP Wave Settings
IUP Wave Setting
Numeric Code
IUP OFF
400
IUP UDEC (uniform deceleration)
401
IUP EDEC (early deceleration)
402
IUP LDEC (late deceleration)
403
IUP UACC (uniform acceleration)
404
To select a contraction period in the Numeric-Control Mode, press the number keys for
one of the settings found in Table 2-34, and select
RUN
. (The selected period is applied
to a run of the specific IUP-wave simulation that was last selected, whether uniform
deceleration, early deceleration, late deceleration, or uniform acceleration.)
Table 2-34. Numeric Codes for IUP Contraction Period Settings
IUP Contraction Period Setting
Numeric Code
IUP MANUAL
405
IUP 2 MIN
406
IUP 3 MIN
407
IUP 5 MIN
408
Содержание Biomedical MPS450 Operators
Страница 6: ......
Страница 13: ...Contents continued vii List of Figures Figure Title Page 1 1 MPS450 Patient Simulator 1 8...
Страница 14: ...MPS450 Operators Manual viii...
Страница 16: ......
Страница 32: ...MPS450 Operators Manual 1 18...
Страница 74: ...MPS450 Operators Manual 2 42...
Страница 76: ...MPS450 Operators Manual 3 2...
Страница 82: ...MPS450 Operators Manual 3 8...