MPS450
Operators Manual
A-2
Table A-1. Problem vs. Solution (cont.)
Problem Probable
Cause
Solution
Respiration
Respiration lead not
selected correctly
Check monitor respiration-lead detection and
switch accordingly on the MPS450.
Leads not connected
properly
Verify that the color code is correct for U.S.
or International ECG leads.
No respiration signal
Performance waveform
currently output (which
turns off respiration)
Select normal physiological waveform.
Temperature
No temperature reading
Wrong series
temperature cable used
Check manufacturer’s series for thermistor
used. Choose the correct Fluke Biomedical
temperature cable for 700 or 400 Series.
Cardiac Output
Selected Injectate
temperature not set
correctly
Injectate temperature must match expected
temperature: 24
°
C or 0
°
C. Use optional
cardiac-output box for setting injectate
temperature.
Incorrect setting for
injectate volume on
cardiac output computer
Set to 10 cc.
Incorrect setting for
catheter size on cardiac
output computer
Set to 7f.
Expected values not correct, or
no values
Correct constant not set
on monitor
Set correct constant to .542 or .595. See
chapter on Cardiac Output.
Injectate temperature not
displayed on monitor
Turn on injectate-temperature pot. (Zero the
box until reading displayed is 0
°
C or 24
°
C.)
Содержание Biomedical MPS450 Operators
Страница 6: ......
Страница 13: ...Contents continued vii List of Figures Figure Title Page 1 1 MPS450 Patient Simulator 1 8...
Страница 14: ...MPS450 Operators Manual viii...
Страница 16: ......
Страница 32: ...MPS450 Operators Manual 1 18...
Страница 74: ...MPS450 Operators Manual 2 42...
Страница 76: ...MPS450 Operators Manual 3 2...
Страница 82: ...MPS450 Operators Manual 3 8...