MPS450
Operators Manual
2-4
3.
The heart-rate setting identified on the LCD screen is active when displayed and
remains active until the setting is changed. (Adjusting the BPM rate does not affect
simulations for arrhythmias, which set their own rates.)
Alternatively, in the Numeric-Control Mode, press the number keys for a heart-rate
setting according to Table 2-1, and select
RUN
:
Table 2-1. Numeric Codes for BPM Settings
BPM Setting
Numeric Code
30 BPM
165
40 BPM
166
45 BPM
250
60 BPM
167
80 BPM
168
90 BPM
251
100 BPM
169
120 BPM
170
140 BPM
171
160 BPM
172
180 BPM
173
200 BPM
174
220 BPM
175
240 BPM
176
260 BPM
177
280 BPM
178
300 BPM
179
Adjusting the ECG Amplitude
Most waveforms are sent to the ECG using the selected adjustment (QRS height) for
amplitude. (ECG amplitudes are used only as a reference during arrhythmia simulations.)
The MPS450 offers a selection of twenty preprogrammed amplitude settings (Lead II):
0.05 to 0.5 mV (0.05 mV steps) and 0.5 to 5.5 mV (0.5 mV steps). This amplitude setting
applies to all ECG waveforms except performance waves and R waves, each of which
has its own amplitude settings.
Action in the Menu-Control Mode
1.
Press the top-menu key labeled NSR. (The top-menu screen for NSR displays
automatically whenever the MPS450 is turned on.)
2.
Select
SEL
to toggle to the screen for adjusting ECG amplitude (bpm/AMPL), with
AMPL in upper-case letters.
3.
To adjust the ECG amplitude, select
DOWN
or
UP
.
4.
The ECG-amplitude setting identified on the LCD screen is active when displayed
and remains active until the setting is changed.
Содержание Biomedical MPS450 Operators
Страница 6: ......
Страница 13: ...Contents continued vii List of Figures Figure Title Page 1 1 MPS450 Patient Simulator 1 8...
Страница 14: ...MPS450 Operators Manual viii...
Страница 16: ......
Страница 32: ...MPS450 Operators Manual 1 18...
Страница 74: ...MPS450 Operators Manual 2 42...
Страница 76: ...MPS450 Operators Manual 3 2...
Страница 82: ...MPS450 Operators Manual 3 8...