Introduction
and
Specifications
Additional Specifications
1
1-23
AC Voltage (Non-Sine Wave) (cont.)
Triangle Wave &
Truncated Sine
Range, p-p
[1]
Frequency
1-Year Absolute Uncertainty,
tcal
±
5
°
C,
±
(% of % of range)
[2]
Max Voltage
Resolution
Auxiliary Output (Dual Output Mode)
0.01 to 10 Hz
5.0 + 0.5
Two digits on each range
10 to 45 Hz
0.25 + 0.5
45 Hz to 1 kHz
0.25 + 0.25
9.3 to 14.0000 V
1 to 10 kHz
5.0 + 0.5
Six digits on each range
[1]
To convert p-p to rms for triangle wave, multiply the p-p value by 0.2886751. To convert p-p to rms for truncated sine wave,
multiply the p-p value by 0.2165063.
[2]
Uncertainty is stated in p-p. Amplitude is verified using an rms-responding DMM.
[3]
Uncertainty for Truncated Sine outputs is typical over this frequency band.
Square Wave
Range
(p-p)
[1]
Frequency
1-Year Absolute Uncertainty,
tcal
±
5
°
C,
±
(% of % of range)
[2]
Max Voltage
Resolution
Normal Channel (Single Output Mode)
0.01 to 10 Hz
5.0 + 0.5
Two digits on each range
10 to 45 Hz
0.25 + 0.5
45 Hz to 1 kHz
0.25 + 0.25
1 to 20 kHz
0.5 + 0.25
2.9 to 65.999 mV
20 to 100 kHz
5.0 + 0.5
Six digits on each range
0.01 to 10 Hz
5.0 + 0.5
Two digits on each range
10 to 45 Hz
0.25 + 0.5
45 Hz to 1 kHz
0.25 + 0.25
1 to 20 kHz
0.5 + 0.25
66 to 659.999 mV
20 to 100 kHz
5.0 + 0.5
Six digits on each range
0.01 to 10 Hz
5.0 + 0.5
Two digits on each range
10 to 45 Hz
0.25 + 0.5
45 Hz to 1 kHz
0.25 + 0.25
1 to 20 kHz
0.5 + 0.25
0.66 to 6.59999 V
20 to 100 kHz
5.0 + 0.5
Six digits on each range
0.01 to 10 Hz
5.0 + 0.5
Two digits on each range
10 to 45 Hz
0.25 + 0.5
45 Hz to 1 kHz
0.25 + 0.25
1 to 20 kHz
0.5 + 0.25
6.6 to 66.0000 V
20 to 100 kHz
5.0 + 0.5
Six digits on each range
Auxiliary Output (Dual Output Mode)
0.01 to 10 Hz
5.0 + 0.5
Two digits on each range
10 to 45 Hz
0.25 + 0.5
45 Hz to 1 kHz
0.25 + 0.25
29 to 659.999 mV
1 to 10 kHz
[3]
5.0 + 0.5
Six digits on each range
0.01 to 10 Hz
5.0 + 0.5
Two digits on each range
10 to 45 Hz
0.25 + 0.5
45 Hz to 1 kHz
0.25 + 0.25
0.66 to 6.59999 V
1 to 10 kHz
[3]
5.0 + 0.5
Six digits on each range
0.01 to 10 Hz
5.0 + 0.5
Two digits on each range
10 to 45 Hz
0.25 + 0.5
45 Hz to 1 kHz
0.25 + 0.25
6.6 to 14.0000 V
1 to 10 kHz
[3]
5.0 + 0.5
Six digits on each range
[1]
To convert p-p to rms for square wave, multiply the p-p value by 0.5.
[2]
Uncertainty is stated in p-p. Amplitude is verified using an rms-responding DMM.
[3]
Limited to 1 kHz for Auxiliary outputs
≥
6.6 V p-p.
Содержание 5522A Operators
Страница 4: ......
Страница 14: ...5522A Operators Manual x ...
Страница 18: ...5522A Operators Manual xiv ...
Страница 22: ...5522A Operators Manual 1 2 ...
Страница 46: ...5522A Operators Manual 1 26 ...
Страница 48: ...5522A Operators Manual 2 2 ...
Страница 54: ...5522A Operators Manual 2 8 ...
Страница 56: ...5522A Operators Manual 3 2 ...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 190: ...5522A Operators Manual 6 2 ...
Страница 240: ...5522A Operators Manual 7 2 ...
Страница 260: ...5522A Operators Manual 8 2 ...
Страница 314: ...5522A Operators Manual 9 52 ...
Страница 370: ...5522A Operators Manual 10 56 ...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 414: ...5522A Operators Manual B 2 ...
Страница 416: ...5522A Operators Manual B 4 ...
Страница 426: ...5522A Operators Manual D 6 ...