Remote Commands
Commands
6
6-37
Returns
PT3926
when a 100-ohm RTD with curve
α
=0.003926 ohms/ohm/
°
C is set as
the RTD type.
RTD_TYPE_D
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Resistance Temperature Detector Type Default command) Sets the default Resistance
Temperature Detector (RTD) at power on and reset, which is saved in the Calibrator non-
volatile memory. (While saving configuration data in the non-volatile memory, a period
of about 2 seconds, the Calibrator does not respond to remote commands.)
Parameters:
PT385
(100-ohm RTD, curve
α
=0.00385 ohms/ohm/
°
C)
PT385_200
(200-ohm RTD, curve
α
=0.00385 ohms/ohm/
°
C)
PT385_500
(500-ohm RTD, curve
α
=0.00385 ohms/ohm/
°
C)
PT385_1000
(1000-ohm RTD, curve
α
=0.00385 ohms/ohm/
°
C)
PT3926
(100-ohm RTD, curve
α
=0.003926 ohms/ohm/
°
C)
PT3916
(100-ohm RTD, curve
α
=0.003916 ohms/ohm/
°
C)
CU10
(10-ohm RTD, empirical curve)
NI120
(120-ohm RTD, empirical curve)
Example:
RTD_TYPE_D PT3926
Set the RTD default type to a 100-ohm RTD with curve
α
=0.003926 ohms/ohm/
°
C.
RTD_TYPE_D?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Resistance Temperature Detector Type Default query) Returns the default Resistance
Temperature Detector (RTD) used at power on and reset.
Responses:
PT385
(100-ohm RTD, curve
α
=0.00385 ohms/ohm/
°
C)
PT385_200
(200-ohm RTD, curve
α
=0.00385 ohms/ohm/
°
C)
PT385_500
(500-ohm RTD, curve
α
=0.00385 ohms/ohm/
°
C)
PT385_1000
(1000-ohm RTD, curve
α
=0.00385 ohms/ohm/
°
C)
PT3926
(100-ohm RTD, curve
α
=0.003926 ohms/ohm/
°
C)
PT3916
(100-ohm RTD, curve
α
=0.003916 ohms/ohm/
°
C)
CU10
(10-ohm RTD, empirical curve)
NI120
(120-ohm RTD, empirical curve)
Example:
RTD_TYPE_D?
returns
PT3926
Returns
PT3926
when the RTD default type is a 100-ohm RTD with curve
α
=0.003926 ohms/ohm/
°
C.
SP_SET
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Host Serial Port Set command) Sets the RS-232-C settings for the Calibrator rear panel
SERIAL 1 FROM HOST serial port, which is saved in the Calibrator non-volatile
memory. (While saving configuration data in the non-volatile memory, a period of about
2 seconds, the Calibrator does not respond to remote commands.) (To set the parameters
for the rear panel SERIAL 2 TO UUT serial port, see the
UUT_SET
command.) The
factory default values are shown below in
bold
type. (To return to the factory defaults,
see the
FORMAT SETUP
command.)
The interface selection sets the command response, with command echo back for
commands and error messages with TERM (terminal) or no echo back with COMP
(computer).
Parameters: <baud rate value>,
300, 600, 1200, 2400, 4800,
9600
Содержание 5522A Operators
Страница 4: ......
Страница 14: ...5522A Operators Manual x ...
Страница 18: ...5522A Operators Manual xiv ...
Страница 22: ...5522A Operators Manual 1 2 ...
Страница 46: ...5522A Operators Manual 1 26 ...
Страница 48: ...5522A Operators Manual 2 2 ...
Страница 54: ...5522A Operators Manual 2 8 ...
Страница 56: ...5522A Operators Manual 3 2 ...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 190: ...5522A Operators Manual 6 2 ...
Страница 240: ...5522A Operators Manual 7 2 ...
Страница 260: ...5522A Operators Manual 8 2 ...
Страница 314: ...5522A Operators Manual 9 52 ...
Страница 370: ...5522A Operators Manual 10 56 ...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 414: ...5522A Operators Manual B 2 ...
Страница 416: ...5522A Operators Manual B 4 ...
Страница 426: ...5522A Operators Manual D 6 ...