5522A
Operators Manual
9-20
duty cycle square wave.) Note that selections available under SHAPE depend on the
selected marker period (frequency) as follows:
Selection Period
(Frequency)
Sine
10 ns – 2 ns (100 MHz – 500 MHz)
Spike
5 s – 20 ns (0.2 Hz – 50 MHz)
Square
5 s – 10 ns (0.2 Hz – 100 MHz)
Sq20%
20 ms – 100 ns (50 kHz – 10 MHz)
•
TRIG
If you are using the external trigger, use this key to cycle through the trigger
settings. The available trigger settings are: off, /1 (trigger signal appears on each
marker), /10 (trigger signal appears on every tenth marker), and /100 (trigger signal
appears on each 100 markers).
•
MODE
Indicates you are in Marker mode. Use the softkey to change modes and
open the corresponding menus for the other four oscilloscope calibration modes.
Default marker values are 1.000 ms, SHAPE = spike.
The
X
and
D
keys step the voltages through cardinal point values of an
oscilloscope in a 1-2-5 step sequence. For example, if the period is 1.000 ms, pressing
X
increases the period to the nearest cardinal point, which is 2.000 ms. Pressing
D
decreases the voltage to the nearest cardinal point, which is 500
μ
s.
Time Base Marker Calibration Procedure for an Oscilloscope
This sample procedure uses the Time Marker function to check the horizontal deflection
(time base) of your oscilloscope. See your oscilloscope’s manual for the exact time base
values recommended for calibration.
Before you begin this procedure, verify that you are in Marker mode. If you are, the
Control Display shows the following menu.
gjh060.eps
Perform the following sample procedure to calibrate the time base.
1.
Connect the calibrator to Channel 1 on the oscilloscope. Select 50
Ω
impedance or
use an external 50
Ω
termination. Make sure the oscilloscope is dc-coupled.
2.
Apply a time marker value according to the recommended calibration settings in your
oscilloscope manual. For example, to enter 200 ns, press
2
0
0
b
K
b
H
, then press
E
.
NOTE
You may enter the equivalent frequency instead of the time marker value.
For example, instead of entering 200 ns, you may enter 5 MHz.
3.
Set your oscilloscope’s time base to show 10 time markers. The time markers should
align with the oscilloscope divisions, as shown in the example below.
For an accurate reading, align the signal’s peaks with the horizontal center axis.
Содержание 5522A Operators
Страница 4: ......
Страница 14: ...5522A Operators Manual x ...
Страница 18: ...5522A Operators Manual xiv ...
Страница 22: ...5522A Operators Manual 1 2 ...
Страница 46: ...5522A Operators Manual 1 26 ...
Страница 48: ...5522A Operators Manual 2 2 ...
Страница 54: ...5522A Operators Manual 2 8 ...
Страница 56: ...5522A Operators Manual 3 2 ...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 190: ...5522A Operators Manual 6 2 ...
Страница 240: ...5522A Operators Manual 7 2 ...
Страница 260: ...5522A Operators Manual 8 2 ...
Страница 314: ...5522A Operators Manual 9 52 ...
Страница 370: ...5522A Operators Manual 10 56 ...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 414: ...5522A Operators Manual B 2 ...
Страница 416: ...5522A Operators Manual B 4 ...
Страница 426: ...5522A Operators Manual D 6 ...