5522A
Operators Manual
7-20
Table 7-15. Verification Tests for Frequency
Range, Normal
Output, V
Output, Normal,
V
Frequency Lower
Limit
[1]
Upper
Limit
[1]
119.00 Hz
118.99970 Hz
119.00030 Hz
120.0 Hz
119.99970 Hz
120.00031 Hz
1000.0 Hz
999.9975 Hz
1000.0025 Hz
3.29999 3.00000
100.00 kHz
99,999.75 Hz
100,000.25 Hz
[1]
Frequency accuracy is specified for 1 year.
Содержание 5522A Operators
Страница 4: ......
Страница 14: ...5522A Operators Manual x ...
Страница 18: ...5522A Operators Manual xiv ...
Страница 22: ...5522A Operators Manual 1 2 ...
Страница 46: ...5522A Operators Manual 1 26 ...
Страница 48: ...5522A Operators Manual 2 2 ...
Страница 54: ...5522A Operators Manual 2 8 ...
Страница 56: ...5522A Operators Manual 3 2 ...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 190: ...5522A Operators Manual 6 2 ...
Страница 240: ...5522A Operators Manual 7 2 ...
Страница 260: ...5522A Operators Manual 8 2 ...
Страница 314: ...5522A Operators Manual 9 52 ...
Страница 370: ...5522A Operators Manual 10 56 ...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 414: ...5522A Operators Manual B 2 ...
Страница 416: ...5522A Operators Manual B 4 ...
Страница 426: ...5522A Operators Manual D 6 ...