Remote Operations
How to Use Commands
5
5-29
You can also use the status commands
*OPC
and
*OPC?
to detect completion of
overlapped commands. (See “Checking 5522A Status.”)
Sequential Commands
Commands that execute immediately are called sequential commands.
In Chapter 6, the command shows a checkbox for sequential commands.
The majority of the commands are sequential.
Commands that Require the Calibration Switch
The following commands do not work unless the rear panel CALIBRATION switch is in
the ENABLE position:
CLOCK
(when setting date but not time)
FORMAT ALL
FORMAT CAL
*PUD
Attempting to use any of these commands with the CALIBRATION switch in the
NORMAL position logs an error into the error queue. (Or it returns the error message if
in the RS-232 Terminal Mode.)
Commands for RS-232 Only
The RS-232 checkbox indicates RS-232 interface commands.
The IEEE-488 and RS-232 interfaces both send commands to the Calibrator as data,
except for those IEEE-488 functions that must be implemented as a message as specified
in the IEEE-488 standards. For example, the RS-232 interface uses the command
REMOTE to place the Calibrator in the remote mode. Although the IEEE-488 interface
could also send a command REMOTE as data, it does not because this is one of the
functions that must be implemented per IEEE-488 Standards. The relationship between
these IEEE-488 messages and the equivalent RS-232 emulation is shown in Table 5-6.
Table 5-6. Commands for RS-232 Only
IEEE-488 Message
[1]
RS-232
Equivalent
GTL LOCAL
command
GTR REMOTE
command
LLO LOCKOUT
command
SRQ SRQSTR
command
SDC, DCL
^C (<Cntl> C) character [clear the device]
GET
^T (<Cntl> T) character [execute a group trigger]
SPE, SPD
^P (<Cntl> P) character [print the serial poll string]
[1]
See “How IEEE-488 Operates” later in this chapter.
In addition to the commands and special characters that emulate the IEEE-488 functions
shown above, there are several more commands that are related to operation and control
of the actual RS-232 Host port and are therefore completely unrelated to IEEE-488
operations. These include the following six commands.
SP_SET
SPLSTR
SRQSTR
SP_SET?
SPLSTR?
SRQSTR?
Содержание 5522A Operators
Страница 4: ......
Страница 14: ...5522A Operators Manual x ...
Страница 18: ...5522A Operators Manual xiv ...
Страница 22: ...5522A Operators Manual 1 2 ...
Страница 46: ...5522A Operators Manual 1 26 ...
Страница 48: ...5522A Operators Manual 2 2 ...
Страница 54: ...5522A Operators Manual 2 8 ...
Страница 56: ...5522A Operators Manual 3 2 ...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 190: ...5522A Operators Manual 6 2 ...
Страница 240: ...5522A Operators Manual 7 2 ...
Страница 260: ...5522A Operators Manual 8 2 ...
Страница 314: ...5522A Operators Manual 9 52 ...
Страница 370: ...5522A Operators Manual 10 56 ...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 414: ...5522A Operators Manual B 2 ...
Страница 416: ...5522A Operators Manual B 4 ...
Страница 426: ...5522A Operators Manual D 6 ...