Remote Commands
Commands
6
6-35
REMOTE
Overlapped
Coupled
IEEE-488
RS-232
Sequential
x
x
x
x
x
(Remote command) Places the Calibrator into the remote state. This command duplicates
the IEEE-488 REN (Remote Enable) message. When in the remote state, the Control
Display shows the softkey “REMOTE CONTROL Go to Local.” Pressing this softkey
returns the Calibrator to local operation If the front panel is locked out, the Control
Display shows the softkey “REMOTE CONTROL LOCAL LOCK OUT.” (See the
LOCKOUT
command.) To unlock the front panel, use the LOCAL command, or cycle the
Calibrator power switch.
Parameter: (None)
Example:
REMOTE
Place the Calibrator in the remote state and display this state on the front panel Control
Display with a softkey REMOTE CONTROL.
RPT_STR
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Report String command) Loads the user report string. The user report string can be read
on the Control Display in local operation, and appears on calibration reports. The
CALIBRATION switch must be set to ENABLE. (Sequential command.)
Parameter:
String of up to 40 characters
RPT_STR?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Report String query) Returns the user report string. The user report string can be read on
the Control Display in local operation, and appears on calibration reports. (Sequential
command.)
Parameter: None
Response:
(String) Up to 40 characters
*RST
Overlapped
Coupled
IEEE-488
RS-232
Sequential
x
x
x
x
x
(Reset Instrument command) Resets the Calibrator to the power-up state. *RST holds off
execution of subsequent commands until the reset operation is complete. This command
is the same as pressing the front panel
R
key.
A reset action evokes the following commands and values:
Command Value Command Value
CUR_POST AUX
REFCLOCK
<REFCLOCK_D
value>
DBMZ
<DBMZ_D value>
REFPHASE
<REFPHASE_D value>
DC_OFFSET 0V
RTD_TYPE
<RTD_TYPE_D
value>
DUTY 50PCT
SCOPE
OFF
EARTH OPEN
STBY
(No
output)
EXTGUARD OFF
TC_OFFSET
0
CEL
HARMONIC 1,
PRI
TC_OTCD
ON
LCOMP OFF
TC_REF
INT
LOWS TIED
TC_TYPE <TC_TYPE_D
value>
OUT 0V,0HZ
TRIG
OFF
Содержание 5522A Operators
Страница 4: ......
Страница 14: ...5522A Operators Manual x ...
Страница 18: ...5522A Operators Manual xiv ...
Страница 22: ...5522A Operators Manual 1 2 ...
Страница 46: ...5522A Operators Manual 1 26 ...
Страница 48: ...5522A Operators Manual 2 2 ...
Страница 54: ...5522A Operators Manual 2 8 ...
Страница 56: ...5522A Operators Manual 3 2 ...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 190: ...5522A Operators Manual 6 2 ...
Страница 240: ...5522A Operators Manual 7 2 ...
Страница 260: ...5522A Operators Manual 8 2 ...
Страница 314: ...5522A Operators Manual 9 52 ...
Страница 370: ...5522A Operators Manual 10 56 ...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 414: ...5522A Operators Manual B 2 ...
Страница 416: ...5522A Operators Manual B 4 ...
Страница 426: ...5522A Operators Manual D 6 ...