5522A
Operators Manual
10-32
RANGE
(IEEE-488, RS-232, Sequential)
Programs the instrument range in PULSE, MEAS Z, OVERLD modes.
Parameters:
Pulse
TP0DB TP8DB TP20DB TP28DB TP40DB TP48DB
Range
2.5 V
1.0 V
250 mV
100 mV
25 mV
10 mV
Impedance
Measure
TZ50OHM TZ1MOHM TZCAP
Range res
50
Ω
res
1M
Ω
cap
Impedance
Measure
TZ50OHM TZ1MOHM TZCAP
Range res
50
Ω
res
1M
Ω
cap
Example:
RANGE TP20DB
Edge Function Commands
TDPULSE
(IEEE-488, RS-232, Sequential)
Turns tunnel diode pulse drive on/off in EDGE mode.
Parameters: ON (or non-zero) or OFF (or zero)
Example:
TDPULSE ON
Returns the tunnel diode pulse drive setting in EDGE mode.
Parameters: None
Response:
1 if ON, 0 if OFF.
Marker Function Commands
TMWAVE
(IEEE-488, RS-232, Sequential)
Selects the waveform for MARKER mode.
Parameters:
SINE
Sine wave (2 ns to 15 ns)
SPIKE
Triangular/sawtooth pulse (15 ns to 5s)
SQUARE
Square wave (50 % duty cycle) (4 ns to 5s)
SQ20PCT
Square wave (20 % duty cycle) (85 ns to 5s)
Example:
TMWAVE SPIKE
Содержание 5522A Operators
Страница 4: ......
Страница 14: ...5522A Operators Manual x ...
Страница 18: ...5522A Operators Manual xiv ...
Страница 22: ...5522A Operators Manual 1 2 ...
Страница 46: ...5522A Operators Manual 1 26 ...
Страница 48: ...5522A Operators Manual 2 2 ...
Страница 54: ...5522A Operators Manual 2 8 ...
Страница 56: ...5522A Operators Manual 3 2 ...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 190: ...5522A Operators Manual 6 2 ...
Страница 240: ...5522A Operators Manual 7 2 ...
Страница 260: ...5522A Operators Manual 8 2 ...
Страница 314: ...5522A Operators Manual 9 52 ...
Страница 370: ...5522A Operators Manual 10 56 ...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 414: ...5522A Operators Manual B 2 ...
Страница 416: ...5522A Operators Manual B 4 ...
Страница 426: ...5522A Operators Manual D 6 ...