5522A
Operators Manual
9-6
Pulse Generator Specifications
Pulse Generator Characteristics
Positive pulse into 50
Ω
Typical rise/fall times
<2 ns
Available Amplitudes
2.5 V, 1 V, 250 mV, 100 mV, 25 mV, 10 mV
Pulse Width
Range
4 ns to 500 ns
[1]
Uncertainty
[2]
5 % of pulse width
±
2 ns
Pulse Period
Range
22 ms to 200 ns (45.5 Hz to 5 MHz)
Resolution
4 or 5 digits depending upon frequency and width
1-Year Absolute Uncertainty at Cardinal Points, tcal
±
5
°
C
±
2.5 ppm
[1]
Pulse width not to exceed 40 % of period.
[2]
Pulse width uncertainties for periods below 2
μ
s are not specified.
Trigger Signal Specifications (Pulse Function)
Pulse Period
Division Ratio
Amplitude into 50
Ω
(p-p)
Typical Rise Time
22 ms to 200 ns
off/1/10/100
≥
1 V
≤
2 ns
Trigger Signal Specifications (Time Marker Function)
Time Marker Period
Division Ratio
Amplitude into 50
Ω
(p-p)
Typical Rise Time
2 to 9 ns
off/100
≥
1 V
≤
2 ns
10 to 749 ns
off/10/100
≥
1 V
≤
2 ns
750 ns to 34.9 ms
off/1/10/100
≥
1 V
≤
2 ns
35 ms to 5 s
off/1
≥
1 V
≤
2 ns
Trigger Signal Specifications (Edge Function)
Edge Signal
Frequency
Division Ratio
Typical Amplitude into
50
Ω
(p-p)
Typical Rise Time
Typical Lead Time
900 Hz to 11 MHz
off/1
≥
1 V
≤
2 ns
40 ns
Trigger Signal Specifications (Square Wave Voltage Function)
Voltage Function
Frequency
Division Ratio
Typical Amplitude into
50
Ω
(p-p)
Typical Rise Time
Typical Lead Time
10 Hz to 10 kHz
off/1
≥
1 V
≤
2 ns
1
μ
s
Trigger Signal Specifications
Trigger Signal Type
Parameters
Field Formats
Selectable NTSC, SECAM, PAL, PAL-M
Polarity
Selectable inverted or uninverted video
Amplitude into 50
Ω
load
Adjustable 0 to 1.5 V p-p
Ω
, (
±
7 % accuracy)
Line Marker
Selectable Line Video Marker
Oscilloscope Input Resistance Measurement Specifications
Scope Input Selected
50
Ω
1
M
Ω
Measurement Range
40 to
60
Ω
500
k
Ω
to
1
.
5 M
Ω
Uncertainty
0.1 %
0.1 %
Oscilloscope Input Capacitance Measurement Specifications
Scope Input selected
1 M
Ω
Measurement Range
5 to 50 pF
Uncertainty
±
(5 % of input + 0.5 pF)
[1]
[1]
Measurement made within 30 minutes of capacitance zero reference. SC600 option must be selected for at least five minutes
prior to any capacitance measurement, including the zero process.
Содержание 5522A Operators
Страница 4: ......
Страница 14: ...5522A Operators Manual x ...
Страница 18: ...5522A Operators Manual xiv ...
Страница 22: ...5522A Operators Manual 1 2 ...
Страница 46: ...5522A Operators Manual 1 26 ...
Страница 48: ...5522A Operators Manual 2 2 ...
Страница 54: ...5522A Operators Manual 2 8 ...
Страница 56: ...5522A Operators Manual 3 2 ...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 190: ...5522A Operators Manual 6 2 ...
Страница 240: ...5522A Operators Manual 7 2 ...
Страница 260: ...5522A Operators Manual 8 2 ...
Страница 314: ...5522A Operators Manual 9 52 ...
Страница 370: ...5522A Operators Manual 10 56 ...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 414: ...5522A Operators Manual B 2 ...
Страница 416: ...5522A Operators Manual B 4 ...
Страница 426: ...5522A Operators Manual D 6 ...