5522A
Operators Manual
6-42
If INT is returned, the reference temperature return is 0 unless you are in a thermocouple
mode of operation and the Calibrator is in Operate.
Responses:
INT
, <value of reference temperature>,
CEL
(or
FAR
)
EXT
, <value of reference temperature>,
CEL
(or
FAR
)
Example:
TC_REF?
returns
INT,2.988E+01,CEL
Returns Internal, 29.88, Celsius, when the thermocouple reference is internal and at
29.88
°
C. (If the temperature return for the internal reference is 0 (
0.00E+00
), the
Calibrator is not in Operate, and/or the Calibrator is not in a thermocouple mode.)
TC_TYPE
Overlapped
Coupled
IEEE-488
RS-232
Sequential
x
x
x
x
x
(Thermocouple Type command) Sets the Thermocouple (TC) temperature sensor type.
The TC type is used when the output is set to a temperature value with the
OUT
command
and the temperature sensor type is set to
TC
with the
TSENS_TYPE
command. When the
thermocouple type is changed while simulating a temperature output, the temperature is
changed to 0
°
C. Once set, the Calibrator retains the TC type until power off or reset.
Parameters:
B
(B-type
thermocouple)
C
(C-type
thermocouple)
E
(E-type
thermocouple)
J
(J-type
thermocouple)
K
(K-type thermocouple) (default)
N
(N-type
thermocouple)
R
(R-type
thermocouple)
S
(S-type
thermocouple)
T
(T-type
thermocouple)
X
(10
μ
V/
°
C linear output)
Y
(% relative humidity)
Z
(1
mV/
°
C linear output)
Example:
TC_TYPE J
Set the thermocouple type for simulating a temperature output to a J-type thermocouple.
TC_TYPE?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Thermocouple Type query) Returns the Thermocouple (TC) temperature sensor type.
When the thermocouple type is changed while simulating a temperature output, the
temperature is changed to 0
°
C.
Responses:
B
(B-type
thermocouple)
C
(C-type
thermocouple)
E
(E-type
thermocouple)
J
(J-type
thermocouple)
K
(K-type thermocouple) (default)
N
(N-type
thermocouple)
R
(R-type
thermocouple)
S
(S-type
thermocouple)
T
(T-type
thermocouple)
X
(10
μ
V/
°
C linear output)
Y
(% relative humidity)
Z
(1
mV/
°
C linear output)
Example:
TC_TYPE?
returns
K
Содержание 5522A Operators
Страница 4: ......
Страница 14: ...5522A Operators Manual x ...
Страница 18: ...5522A Operators Manual xiv ...
Страница 22: ...5522A Operators Manual 1 2 ...
Страница 46: ...5522A Operators Manual 1 26 ...
Страница 48: ...5522A Operators Manual 2 2 ...
Страница 54: ...5522A Operators Manual 2 8 ...
Страница 56: ...5522A Operators Manual 3 2 ...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 190: ...5522A Operators Manual 6 2 ...
Страница 240: ...5522A Operators Manual 7 2 ...
Страница 260: ...5522A Operators Manual 8 2 ...
Страница 314: ...5522A Operators Manual 9 52 ...
Страница 370: ...5522A Operators Manual 10 56 ...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 414: ...5522A Operators Manual B 2 ...
Страница 416: ...5522A Operators Manual B 4 ...
Страница 426: ...5522A Operators Manual D 6 ...