5522A
Operators Manual
10-36
Table 10-2. SC1100 Option DC Voltage Verification (cont.)
Nominal
Value (V dc)
Measured Value
(V dc)
Deviation
(V dc)
1-Year Spec. (V dc)
0.499
0.0002895
-0.499
0.0002895
0.5
0.00029
-0.5
0.00029
1.35
0.000715
-1.35
0.000715
2.19
0.001135
-2.19
0.001135
2.2
0.00114
-2.2
0.00114
6.6
0.00334
-6.6
0.00334
10.99
0.005535
-10.99
0.005535
11
0.00554
-11
0.00554
70.5
0.03529
-70.5
0.03529
130
0.06504
-130
0.06504
Table 10-3. SC1100 Option DC Voltage Verification at 50
Ω
Calibrator
Mainframe
Output
Measured Value
(V DC)
Reading x correction
Tolerance (V DC)
0 mV
0.00004 V
2.49 mV
4.623E-05 V
-2.49 mV
4.623E-05 V
9.90 mV
6.475E-05 V
-9.90 mV
6.475E-05 V
24.9 mV
0.0001023 V
-24.9 mV
0.0001023 V
109.9 mV
0.0003148 V
Содержание 5522A Operators
Страница 4: ......
Страница 14: ...5522A Operators Manual x ...
Страница 18: ...5522A Operators Manual xiv ...
Страница 22: ...5522A Operators Manual 1 2 ...
Страница 46: ...5522A Operators Manual 1 26 ...
Страница 48: ...5522A Operators Manual 2 2 ...
Страница 54: ...5522A Operators Manual 2 8 ...
Страница 56: ...5522A Operators Manual 3 2 ...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 190: ...5522A Operators Manual 6 2 ...
Страница 240: ...5522A Operators Manual 7 2 ...
Страница 260: ...5522A Operators Manual 8 2 ...
Страница 314: ...5522A Operators Manual 9 52 ...
Страница 370: ...5522A Operators Manual 10 56 ...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 414: ...5522A Operators Manual B 2 ...
Страница 416: ...5522A Operators Manual B 4 ...
Страница 426: ...5522A Operators Manual D 6 ...