5522A
Operators Manual
10-4
Calibration (tcal).................................................. 15
°
C to 35
°
C
Storage ............................................................... -20
°
C to +70
°
C
Electromagnetic Compatibility
............................ Designed to operate in Standard Laboratory environments where the
Electromagnetic environment is highly controlled. If used in areas with
Electromagnetic fields >1 V/m, there could be errors in output values.
All testing for this specification used new cables and connectors.
Temperature Coefficient
....................................... Temperature Coefficient for temperatures outside tcal +5
°
C is
10 % per
°
C of 1-year specification.
Relative Humidity
Operating ............................................................ <80 % to 30
°
C, <70 % to 40
°
C,<40 % to 50
°
C
Storage ............................................................... <95 %, noncondensing
Altitude
Operating ............................................................ 3,050 m (10,000 ft) maximum
Nonoperating ...................................................... 12,200 m (40,000 ft) maximum
Safety
..................................................................... Designed to comply with IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994;
CAN/CSA-C22.2 No. 1010.1-92
Analog Low Isolation
............................................ 20 V
EMC
........................................................................ Complies with EN 61326-1/1997, Class A
Volt Specifications
DC Signal
Square Wave Signal
[1]
Volt Function
50
Ω
Load
1 M
Ω
Load
50
Ω
Load
1 M
Ω
Load
Amplitude Characteristics
Range
0 to
±
6.6 V
0 to
±
130 V
±
1 mV to
±
6.6 V p-p
±
1 mV to
±
130 V p-p
Resolution
Range
1 to 24.999 mV
25 to 109.99 mV
110 mV to 2.1999 V
2.2 to 10.999 V
11 to 130 V
Resolution
1
μ
V
10
μ
V
100
μ
V
1 mV
10 mV
Adjustment Range
Continuously adjustable
1-Year Absolute Uncertainty, tcal
±
5
°
C
±
(0.25 % of
40
μ
V)
±
(0.05 % of
40
μ
V)
±
(0.25 % of
40
μ
V)
±
(0.1% of
40
μ
V)
[2]
Sequence
1-2-5 (e.g., 10 mV, 20 mV, 50 mV)
Square Wave Frequency Characteristics
Range
10 Hz to 10 kHz
1-Year Absolute Uncertainty, tcal
±
5
°
C
±
(2.5 ppm of setting)
Typical Aberration
within 4
μ
s from 50 % of
leading/trailing edge
<(0.5 % of 100
μ
V)
[1]
Selectable positive or negative, zero referenced square wave.
[2]
For square wave frequencies above 1 kHz,
±
(0.25 % of 40
μ
V).
Содержание 5522A Operators
Страница 4: ......
Страница 14: ...5522A Operators Manual x ...
Страница 18: ...5522A Operators Manual xiv ...
Страница 22: ...5522A Operators Manual 1 2 ...
Страница 46: ...5522A Operators Manual 1 26 ...
Страница 48: ...5522A Operators Manual 2 2 ...
Страница 54: ...5522A Operators Manual 2 8 ...
Страница 56: ...5522A Operators Manual 3 2 ...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 190: ...5522A Operators Manual 6 2 ...
Страница 240: ...5522A Operators Manual 7 2 ...
Страница 260: ...5522A Operators Manual 8 2 ...