Remote Commands
Commands
6
6-15
EARTH
Overlapped
Coupled
IEEE-488
RS-232
Sequential
x
x
x
x
x
(Earth Ground command) Selects whether or not the Calibrator front panel NORMAL
LO terminal is tied to chassis (earth) ground. Once set, the Calibrator retains the earth
setting until power off or reset.
Parameters:
OPEN
(disconnect front panel LO terminal from chassis ground)
TIED
(connect front panel LO terminal to chassis ground)
Example:
EARTH TIED
Load
TIED
to tie the Calibrator front panel NORMAL LO terminal to earth (the front
panel
Z
key annunciator is on).
EARTH?
Overlapped
Coupled
IEEE-488
RS-232
Sequential
x
x
x
x
x
(Earth Ground query) Returns whether or not the Calibrator front panel NORMAL LO
terminal is tied to chassis (earth) ground.
Responses: (character)
OPEN
(front panel LO terminal disconnected from chassis
ground)
(character)
TIED
(front panel LO terminal connected to chassis ground)
Example:
EARTH?
returns
OPEN
Returns
OPEN
when EARTH is not tied to the NORMAL LO terminal (the front panel
Z
key annunciator is off).
EDIT
Overlapped
Coupled
IEEE-488
RS-232
Sequential
x
x
x
x
x
(Edit command) Sets the edit field to the primary, secondary or frequency field.
Parameters:
PRI
(edit the value in single output functions and the primary output
value in dual output functions)
SEC
(edit the secondary value in dual output functions)
FREQ
(edit the frequency value in single ac output functions)
OFF
(edit is off, which is the same as using the NEWREF command)
Example:
EDIT FREQ
Load
FREQ
into the edit field to edit frequency.
EDIT?
Overlapped
Coupled
IEEE-488
RS-232
Sequential
x
x
x
x
x
(Edit query) Returns the edit field setting.
Responses: (character)
PRI
(value in single output functions, and the primary output
value in dual output functions is in edit)
(character)
SEC
(secondary value in dual output functions is in edit)
(character)
FREQ
(frequency value in single ac output functions is in edit)
(character)
OFF
(no value is in edit.)
Example:
EDIT?
returns
OFF
Returns
OFF
when no value is in edit.
Содержание 5522A Operators
Страница 4: ......
Страница 14: ...5522A Operators Manual x ...
Страница 18: ...5522A Operators Manual xiv ...
Страница 22: ...5522A Operators Manual 1 2 ...
Страница 46: ...5522A Operators Manual 1 26 ...
Страница 48: ...5522A Operators Manual 2 2 ...
Страница 54: ...5522A Operators Manual 2 8 ...
Страница 56: ...5522A Operators Manual 3 2 ...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 190: ...5522A Operators Manual 6 2 ...
Страница 240: ...5522A Operators Manual 7 2 ...
Страница 260: ...5522A Operators Manual 8 2 ...
Страница 314: ...5522A Operators Manual 9 52 ...
Страница 370: ...5522A Operators Manual 10 56 ...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 414: ...5522A Operators Manual B 2 ...
Страница 416: ...5522A Operators Manual B 4 ...
Страница 426: ...5522A Operators Manual D 6 ...