5 Operation
StamoSens CSM 750 / CSS 70
16
EHauser
5.2
Local operation
5.2.1
Main menu
Access the main menu by holding down the
H
key, until "MEASUREMENT" is displayed.
The main menu options contain the following fields:
Option
LC display
Info
MEASUREMENT
• Acquiring and displaying
– the sensor signal
– the analog current
– the sensor frequency
• Setting the alarm value switches
• Displaying error messages
PARAMETER ENTRY
Setting the default values for
– Measuring ranges
– Alarm values
– Cleaning
CALIBRATION
POINTS
Defining the number of calibration points
SAK INPUT
Allocating the concentration values to the
calibration measuring points
FREQUENCY
Allocating the frequencies to the measuring points
from the calibration
CONFIGURATION
Setting the default values for
– Sensor type
– Unit of measure
– Correction factor
– Analog output
– Alarm values
– Fault message
– Date, time
LANGUAGE
Selecting menu language
DIAGNOSTICS
Displaying error messages
SERVICE
Manual measurement using a flash for service and
test purposes
DATA MEMORY
Last 340 measured values with the time of their
storage.
Only for service purposes.