6 Commissioning
StamoSens CSM 750 / CSS 70
24
EHauser
Hold down the
W
and
V
keys simultaneously to activate the default setup
for the selected sensor type. The default setup is compulsory for initial start-
up, in order to use the correct default calibration.
For a summary of the sensor’s default setup, see
→
Chap. 11.2
Confirm the "Correction factor 0%" function with the
F
key
Use the
W
or
V
keys to select the number of measured values for forming
the mean value.
Confirm with the
F
key
Use the
W
or
V
keys to select the current output.
Confirm with the
F
key
Use the
W
or
V
keys to select the configuration for alarm value A.
Confirm with the
F
key
Use the
W
or
V
keys to select the configuration for alarm value B.
Confirm with the
F
key
Use the
W
or
V
keys to select the configuration for the diagnostic alarms.
Confirm with the
F
key
Use the
W
,
V
and
F
keys to set the current date and time.
Confirm with the
F
key
Press the
H
key.
The system is now ready for measurement.
Key
Display
n mean value
10
Analog output
4-20 mA
Alarm B
normally open
Diagnostic alarm
normally open
act. Date/Time
10.02.02 11:32