StamoSens CSM 750 / CSS 70
4 Electrical connection
Hauser
13
Fig. 8:
Cleaning unit connection
4.4
Serial interface
!
Note!
The serial interface is only available for service purposes.
Configure the PC as follows:
•
Software protocol: 9600 Baud, N (= no parity), 8 Data bits, 1 Stop bit
•
Output format: ASCII. For a data transfer, the PC must send the ASCII code "81" (hex
51, "Q" = Question).
RS 232-0 of CSM 750 Transmitter
COM 1 / 2 on PC
Terminal
SUB-D
9-pin
Function
Function
SUB-D
9-pin
SUB-D
25-pin
23
3
TxD
RxD
2
3
22
2
RxD
TxD
3
2
21
8
CTS
RTS
7
4
CTS
8
5
20
5
GND
GND
5
7
Содержание Stamosens CSM750