5 Operation
StamoSens CSM 750 / CSS 70
22
EHauser
5.2.10
"SERVICE" Menu
5.2.11
"DATA MEMORY" Menu
!
Note!
"DATA MEMORY" menu can only be read out for service purposes.
Error messages
–
Error messages and trouble-shooting s. Chap. 8.1
Option
Display
Range of adjustment
Factory settings (bold)
Info
Option
Display
Range of adjustment
Factory settings (bold)
Info
SERVICE
Manual measurement using a flash for service and test
purposes
Single measurement
Frequency of the
flash
–
Press
J
to trigger the single measurement. The frequency
measured is displayed on the LC display. The frequency
displayed is meant as a function verification. It can not be
stored in the data memory.
Use the
V
or
W
keys to switch the cleaning contact on or
off.
Option
Display
Range of adjustment
Factory settings (bold)
Info
DATA MEMORY
Last 340 measured values with the time of their storage
Measured value
Moment of
measurement
–
Data sets can only be read
Serial output
yes
no
By pressing the
J
key you can send the serial output of
the data sets via the RS 232 interface in ASCII format. PC
configuration
→
Chap. 4.4
Clear data
yes
no
This deletes all data sets in the data memory
Browse through the data sets by pressing the
W
or
V
keys
–
If the data memory is empty, adjacent displays appear.