StamoSens CSM 750 / CSS 70
10 Technical data
Hauser
35
10
Technical data
10.1
CSM 750
Transmitter
10.1.1
Input
10.1.2
Output
10.1.3
Power supply
Measured variable
Dissolved organic substance
Unit of measure
1/m SAK; mg/l DOC; mg/l TOC; mg/l BOD; mg/l COD
Measuring range
SAK 0-50
0 ... 50 1/m SAK; 0 ... 80 mg/l DOC;
0 ...80 mg/l TOC; 0 ... 80 mg/l BOD;
0 ... 80 mg/l COD
traceable to caliumhydrogenphtalate standard
SAK 0-700
0 ... 700 1/m SAK; 0 ... 900 mg/l DOC;
0 ... 900 mg/l TOC; 0 ... 900 mg/l BOD;
0 ... 900 mg/l COD
traceable to caliumhydrogenphtalate standard
Measuring frequency
0 ... 5965 Hz
Current output
0 ... 20 mA, electrically isolated
4 ... 20 mA, electrically isolated
Load
max. 500
Ω
Relay outputs
2 limit contactors, fault message, hold function, sensor cleaning
Switching capacity of the
contacts
230 V AC / 3 A; 30 V DC / 1 A
Data interface
RS 232
Supply voltage
80 ... 250 V AC ±10%, 50/60 Hz
24 V AC / DC
Power consumption
max. 15 VA
Current consumption
80 ... 250 V AC:
max. 0.2 A
24 V AC / DC:
max. 0.7 A
Fuse
80 ... 250 V AC:
5 x 20 mm, time-lag, 500 mA fine-wire fuse
24 V AC / DC:
5 x 20 mm, time-lag, 2 A fine-wire fuse
Содержание Stamosens CSM750
Страница 2: ...StamoSens CSM 750 CSS 70 2 Endress Hauser...
Страница 38: ...11 Appendix StamoSens CSM 750 CSS 70 38 Endress Hauser 11 Appendix 11 1 Operating matrix...
Страница 39: ...StamoSens CSM 750 CSS 70 11 Appendix Endress Hauser 39...
Страница 46: ...StamoSens CSM 750 CSS 70 46 Endress Hauser...