5 Operation
Hauser
21
5.2.8
"LANGUAGE" Menu
5.2.9
"DIAGNOSTICS" Menu
Alarm value A
NC current
NO current
Setting the relay as NO or NC contact
Changes to this setting will only be activated after
switching the
CSM 750 transmitter on and off.
Alarm value B
NC current
NO current
Diagnostic alarm
NC current
NO current
Date / time
01.01.00 -31.12.99
00:00 - 23:59
current date and time
Factory setting displays current date and time.
Option
Display
Range of adjustment
Factory settings (bold)
Info
Alarm B
normally open
Diagnostic alarm
normally open
act. Date/Time
10.02.02 11:32
Option
Display
Range of adjustment
Factory settings (bold)
Info
LANGUAGE
Selecting menu language
Language
Deutsch
English
Français
Svenska
Suomi
Nederlands
Italiano
Espanol
Polski
Other languages than those provided in the device, on
request
Option
Display
Range of adjustment
Factory settings (bold)
Info
LED display
In the event of a malfunction or error situation, adjacent displays
appear.
Then go into the DIAGNOSTICS menu to access the error messages.
!
Note
• The signal output is only set to "fault" when the cause of a fault is present for an
uninterrupted period of min. 10 s.
• The current output holds the last measured value for the duration of the error. The
limit value alarms remain unchanged.
LC display
DIAGNOSTICS
Displaying error messages