StamoSens CSM 750 / CSS 70
6 Commissioning
Hauser
23
6
Commissioning
6.1
Function check
#
Warning!
Before switching the system on, ensure that the measuring point is not at risk.
Uncontrolled pumps, valves or similar things can lead to devices being damaged.
"
Caution!
•
Before switching on the system check all the connections again for correctness.
•
Also ensure that the post-connection check (s. Chap. 4.5) has been carried out.
6.2
Initial start-up
!
Note!
•
Before switching the system on for the first time, familiarise yourself with the operation
of the transmitter. For this, see in particular Chapters 1 (Safety instructions) and
5 (Operation).
•
The CSS 70 Sensor should be left in the medium for approx. 1 hour, so that the sensor
can adapt to the medium temperature.
•
The CSM 750 Transmitter has already been pre-configured at the factory and starts
measuring automatically when switched on. Due to the different waste water
composition, the media can vary greatly. Therefore, we recommend using an
application-specific calibration for commissioning.
For initial start-up of the measuring system, proceed as follows:
Key
Display
Switch the transmitter on.
Press the
H
key
Press the
W
key 5 times
Confirm with the
F
key
Press the
W
key to set the "99" code.
Confirm with the
F
key
Use the
W
or
V
keys to select your sensor.
Confirm with the
F
key
Use the
W
or
V
keys to select the desired unit of measure.
(1/m SAK, mg/l DOC, mg/l TOC, mg/l BOD, mg/l COD)
Confirm with the
F
key