StamoSens CSM 750 / CSS 70
5 Operation
Hauser
19
5.2.4
"CALIBRATION POINTS" Menu
5.2.5
"SAK INPUT" Menu
5.2.6
"FREQUENCY" Menu
Option
Display
Range of adjustment
Factory settings (bold)
Info
CALIBRATION
POINTS
Defining the number of calibration points
Code number
0 ... 99
0
To access the sub-menu, enter the code "99". If the
incorrect code is entered, the program exits to the
"MEASUREMENT" main menu.
Number of
measuring points
1 ... 7
1
Number of calibration points that should be used to form
the calibration characteristic curve. The device is pre-
calibrated before shipment.
Option
Display
Range of adjustment
Factory settings (bold)
Info
SAK INPUT
Allocating the concentration values to the calibration
measuring points
Code number
0 ... 99
0
To access the sub-menu, enter the code "99". If the
incorrect code is entered, the program exits to the
"MEASUREMENT" main menu.
Enter the
measuring point
1 ... 7
1
Allocating the concentration values from the calibration for
the subsequent frequency input.
Option
Display
Range of adjustment
Factory settings (bold)
Info
FREQUENCY
Allocating the frequencies to the measuring points from the
calibration
Code number
0 ... 99
0
To access the sub-menu, enter the code "99". If the
incorrect code is entered, the program exits to the
"MEASUREMENT" main menu.
FREQUENCY
0 ... 5965 Hz
5312
Entering the measuring frequency for the measuring points
from the calibration.
Factory settings for the measuring frequency depending on
sensor type selected in the CONFIGURATION menu