5 Operation
StamoSens CSM 750 / CSS 70
20
EHauser
5.2.7
"CONFIGURATION" Menu
!
Note!
Some settings that can be made in this menu affect the defaults in the "PARAMETER
ENTRY" menu. In view of this, complete the "CONFIGURATION" menu first during initial
start-up (s. Chap. 6.2).
Option
Display
Range of adjustment
Factory settings (bold)
Info
CONFIGURATION
Setting the default values for
– Sensor type
– Unit of measure
– Correction factor
– Analog output
– Alarm values
– Fault message
– Date, time
Code number
0 ... 99
0
To access the sub-menu, enter the code "99". If the
incorrect code is entered, the program exits to the
"MEASUREMENT" main menu.
Sensor type
SAK 0-50
SAK 0-700
NOx AS 20-50*
NOx AS 0-25*
NOx CW 20-50*
NOx CW 0-25*
NOx AS 0-30*
NOx CW 0-60*
Selection of your chosen CSS 70 Sensor type.
Selection of the sensor type is checked for plausibility, i.e.
whether the selected sensor is identical to the connected
sensor.
Those sensor types marked with * are only available with
the CNS 70 Sensor.
Unit of measure
1/m SAK
mg/l DOC
mg/l TOC
mg/l BOD
mg/l COD
mg/l*
mg/l-N*
Selection of unit of measure.
Those units of measure marked with * are only available for
CNS 70 Sensor operation.
Default setup
yes
no
Selecting "Default setup" loads the sensor default
parameters. With this, the sensor is immediately ready for
measurement, without having to carry out an adaptation.
For a summary of the sensor's factory settings, see
→
Chap. 11.2.
!
Note: If you load the Factory Setting, the previous
values stored are overwritten.
Correction factor
-25...+25%
0%
Linear correction factor of calibration curve at low
fluctuation of waste water composition. Automatically set to
"0" when factory setting is selected.
Mean value
1-20
10
Number of individual measured values that are used for
calculating the arithmetic mean before the measured value
is displayed. This function is used for smoothing the daily
curve.
Analog output
0-20 mA
4-20 mA
Selection of the current range, on which the measuring
range should be mapped.
n mean value
10
Analog output
4-20 mA