12.1 Handheld device
• Height: 36 mm
• Width: 64 mm
• Length: 150 mm
• Weight: (with battery) approx. 437 g
• Power demand:
• Power down: 35 µA
• Idle:
- Background lighting off: 26 mA
- Background lighting on: 56 mA
• Probe switched on: 100 mA
• Measurement: 350 mA
• Measurements per charge: up to approx. 5000 (20 °C / background lighting max.)
• Probes that can be connected: SWZ, S1, S1C, S2
• Storage temperature: –20 to +70 °C (–4 to +158 °F)
• Operating temperature: –20 to +70 °C (–4 to +158 °F)
• Charging temperature: 10 to 30 °C (50 to 86 °F)
• Charging voltage: nom. 12 V, max. 15 V, min. 12 V
• Charging current: approx. 1 A
• Charging time: approx. 2 hours if battery fully discharged
• Accumulator: Ni-MH (4 × 1.2 V) (AA), 2
000 mA/h, >1000 measurements
• Physical BUS: RS485
• Bus protocol: IMP-BUS protocol II
12.2 SWZ probe
• Power supply: 12 to 24 V
DC
• Current consumption: 150 mA @ 12 V
DC
during 2 to 3 s measuring cycle time
• Measuring range: 0 to 100 % vol. water content
• Repeatability, water content measurement (with probe at rest in concrete): ±2 l/m³
• Absolute accuracy: ±3 % of the water quantity
• Conductivity range: 0 to 20 dS/m
• Measuring volume: 0.5 l
• Probe temperature range: 0 to 50 °C (32 to 122 °F)
• Calibration:
• Pre-programmed calibrations for fresh concrete
• Own calibrations possible
• Up to 15 calibration curves can be saved in memory
• Degree of protection: IP68
• Dimensions: 155 mm× 60 mm
• Interfaces: 1.5 m cable with 7-pin coupling socket
12.3 S1 probe
For moisture measurement of bulk solids such as sand and gravel
• Sensor with integrated TDR electronics
• Screw thread: M28 × 1.5 (on cable side)