Solitrend MMP20 (Option D)
S1 moisture probe
Hauser
45
higher-value polynomial is useful to achieve better accuracy, the linear equation often
suffices, particularly in the lower moisture range, and delivers very good results.
Two material samples with different moisture values are required to perform a 2-
point material calibration. The moisture values must be determined with another
laboratory method (e.g. moisture analyzer, kiln drying) before the calibration. It is
important to adhere to the following sequence: first "lower moisture value" (drier
material) and then "upper moisture value" (wetter material).
1. Set the calibration memory (01 – 15) to be overwritten value using the navigation
keys
2. Press the Enter key
The setting is accepted
3. Select the percentage moisture of the lower moisture value using the navigation keys
4. Press the Enter key
The setting is accepted
5. Press the Enter key again
The material measurement process is started
Four measurements are taken in order to increase the accuracy. An average of
these measurement values is then calculated. The measuring time is
approximately 20 seconds. When the measurement is finished, the measured
pulse transit time is displayed briefly.
6. Select the percentage moisture of the upper moisture value using the navigation keys
7. Press the Enter key
The setting is accepted