Solitrend MMP20 (Option D)
S1 moisture probe
Hauser
49
Take the measurements according to the following procedure:
1. Fill the container with sand
A0040894
2. Lift the container approximately 5 cm and then drop it. Repeat this five times (more
if necessary).
This packs (compresses) the sand.
=
5x
A0040895
3. Insert the probe into the sand. Once the base of the probe reaches the sand surface
push the probe in slightly further (do not jiggle the probe or turn it while inserting
it!). In the case of gravel and grit, shake the container while introducing the probe.
Otherwise it is very difficult to introduce the probe into the material. Shaking the
container places the material optimally around the probe rods.
A0040896
A0040897