S1 moisture probe
Solitrend MMP20 (Option D)
50
Hauser
4. Take the measurement with the handheld device
5. Remove the probe from the sand and shake the sand to loosen it again
6. Repeat steps 2 to 4 two more times so that you have 3 measured values in total
7. Pour the sand into a second bucket so you can take readings from the sand at the
bottom (this is particularly relevant for gravel and if the sand is close to saturation as
any free water can settle at the bottom of the container!)
A0040908
8. Repeat steps 2 to 4 three more times so that you have 6 measured values in the end
9. Document the average of the 6 measurements
11.4.3 Replacing the probe rods
The probe rods in the S1 probe can only be replaced by the Service Department.
Содержание Solitrend MMP20
Страница 55: ......
Страница 56: ...www addresses endress com 71462899 71462899...