Solitrend MMP20 (Option D)
S1 moisture probe
Hauser
43
2. Press the Enter key
This activates the selected calibration
Afterwards, the symbol appears in the top right-hand corner of the display
indicating that the option has been activated. In addition, the "!" symbol appears
in front of the active calibration.
Press the navigation key to go directly from the measuring screen to the "Choose"
menu item
"Choose" menu item
You can perform your own material calibrations or customize existing calibrations to suit
your particular needs. Two options are available for this:
•
1-point calibration:
• Adjusts the calibration curve to the selected point
• The slope is not changed as a result
• Just one material sample is required
•
2-point calibration:
• Creation of a linear calibration between two measured points
• Two material samples with different moisture values are required
1-point calibration:
This material calibration option merely adjusts (offsets) the configured calibration. As the
slope is not changed, it is important at the start to select a calibration curve that suits the
material.
A sample of the material being measured is required to perform a 1-point material
calibration. The moisture value must be determined with another laboratory method
(e.g. moisture analyzer, kiln drying) before the calibration.
1. Set the calibration memory (01 – 15) to be overwritten value using the navigation
keys
2. Press the Enter key
The setting is accepted