S1 moisture probe
Solitrend MMP20 (Option D)
44
Hauser
3. Select the percentage moisture using the navigation keys
4. Press the Enter key
The setting is accepted
5. Press the Enter key again
The material measurement process is started
Four measurements are taken in order to increase the accuracy. An average of
these measurement values is then calculated. The measuring time is
approximately 20 seconds. When the measurement is finished, the measured
pulse transit time is displayed briefly.
6. After this, the calibration can be saved to the calibration memory set at the start
("Save").
7. Press the Enter key
The selected memory is overwritten.
The word "OWN:" now appears in front of the original memory name to clearly
indicate which memory has been overwritten.
NOTICE
If "SAVE" is selected at the end of the calibration, one of the pre-configured (or already
modified) calibrations in the probe is overwritten!
‣
The original calibrations can only be restored by our Service Department.
Before the measurement commences, make sure that the rods of the probe are fully
immersed in the material to be measured. The probe must remain in the material for
the entire duration of the measurement and should not be moved.
2-point calibration:
In a 2-point calibration, two material samples of different moisture content are measured
and a linear equation (f(x)=mx+b) is then calculated from this information. While a