5 Operation
Proline Prowirl 72
32
E Hauser
5.4.2
Device variables and process variables
Device variables:
The following device variables are available via the HART protocol:
Process variables:
At the factory, the process variables are assigned to the following device variables:
• Primary process variable (PV)
→
flow
• Secondary process variable (SV)
→
totalizer
• Third process variable (TV)
→
not assigned
• Fourth process variable (FV)
→
not assigned
5.4.3
Universal / common practice HART commands
The following table contains all the universal and common practice commands supported by the
measuring device.
ID (decimal)
Device variable
0
OFF (not assigned)
1
Flow
250
Totalizer
Command no.
HART command / access type
Command data
(numeric data in decimal form)
Response data
(numeric data in decimal form)
Universal commands
0
Read the unique device
identifier
Access type = Read
None
The device identifier provides information on the device and
manufacturer; it cannot be altered.
The response consists of a 12-byte device ID:
– Byte 0: fixed value 254
– Byte 1: manufacturer ID, 17 = E+H
– Byte 2: device type ID, 56 = Prowirl 72
– Byte 3: number of preambles
– Byte 4: rev. no. universal commands
– Byte 5: rev. no. device-spec. Commands
– Byte 6: software revision
– Byte 7: hardware revision
– Byte 8: additional device information
– Byte 9-11: device identification
1
Read the primary process
variable
Access type = Read
None
– Byte 0: HART unit ID of the primary process variable
– Byte 1-4: primary process variable
Primary process variable = flow
!
Note!
Manufacturer-specific units are represented using the HART unit
ID “240”.
2
Read the primary process
variable as current in mA and
percentage of the set
measuring range
Access type = Read
None
– Byte 0-3: current current of the primary process variable in mA
– Byte 4-7: percentage of the set measuring range
Primary process variable = flow
Содержание Proline Prowirl 72
Страница 4: ...QUICK SETUP for quick commissioning Proline Prowirl 72 4 Endress Hauser...
Страница 20: ...3 Installation Proline Prowirl 72 20 Endress Hauser...
Страница 42: ...6 Commissioning Proline Prowirl 72 42 Endress Hauser...
Страница 44: ...7 Maintenance Proline Prowirl 72 44 Endress Hauser...
Страница 114: ...www endress com worldwide BA084D 06 en 12 05 71008404 FM SGML 6 0...