Proline Prowirl 72
11 Description of device functions
E Hauser
103
11.2.15 Group SIMULATION SYSTEM
OPERATION HOURS
(in preparation)
The hours of operation of the device appear on the display.
Display:
Depends on the number of hours of operation elapsed:
Hours of operation < 10 hours
→
display format = 0:00:00 (hr:min:sec)
Hours of operation 10...10,000 hours
→
display format = 0000:00 (hr:min)
Hours of operation < 10,000 hours
→
display format = 000000 (hr)
Function description SIMULATION SYSTEM
SIMULATION FAILSAFE
MODE
Use this function to set all inputs, outputs and the totalizer to their error-response modes,
in order to check whether they respond correctly. During this time, the message #691
“SIMULATION FAILSAFE” appears on the display.
Options:
OFF
ON
Factory setting:
OFF
SIMULATION
MEASURAND
Use this function to set all inputs, outputs and the totalizer to their defined flow-response
modes, in order to check whether they respond correctly.
During this time, the message “#692 SIMULATION MEASURAND” appears on the
display.
Options:
OFF
FLOW
Factory setting:
OFF
"
Caution!
• The measuring device can only be used for measuring to a certain extent while the
simulation is in progress.
• The setting is not saved if the power supply fails.
VALUE SIMULATION
MEASURAND
!
Note!
This function is not available unless the SIMULATION MEASURAND function is active.
Use this function to specify a selectable value (e.g. 12 dm³/s). This value is used to test
downstream devices and the measuring device itself.
User input:
5-digit floating-point number
Factory setting:
0
!
Note!
The appropriate unit is taken from the UNIT FLOW function (see Page 74).
"
Caution!
The setting is not saved if the power supply fails.
Function description SUPERVISION
Содержание Proline Prowirl 72
Страница 4: ...QUICK SETUP for quick commissioning Proline Prowirl 72 4 Endress Hauser...
Страница 20: ...3 Installation Proline Prowirl 72 20 Endress Hauser...
Страница 42: ...6 Commissioning Proline Prowirl 72 42 Endress Hauser...
Страница 44: ...7 Maintenance Proline Prowirl 72 44 Endress Hauser...
Страница 114: ...www endress com worldwide BA084D 06 en 12 05 71008404 FM SGML 6 0...