Proline Prowirl 72
11 Description of device functions
E Hauser
101
AMPLIFICATION
Devices are always optimally configured for the process conditions you specified.
Under certain process conditions, however, interference signals (e.g. strong vibrations)
can be suppressed or the measuring range extended by adjusting the amplification.
The amplification is configured as follows:
• A larger value can be entered for the amplification if the fluid is slow-flowing, the
density is low and there are minor disturbance influences (e.g. plant vibrations).
• A smaller value can be entered for the amplification if the fluid is fast-flowing, the
density is high and there are strong disturbance influences (e.g. plant vibrations).
"
Caution!
Incorrectly configured amplification can have the following effects:
• The measuring range is limited in such a way that small flows cannot be recorded or
displayed. In this instance, the value for the amplification must be increased.
• Undesired interference signals are registered by the device which means that a flow is
recorded and displayed even if the fluid is at a standstill. In this instance, the value for
the amplification must be reduced.
Options:
1...5 (1 = smallest amplification, 5 = largest amplification)
Factory setting:
3
Function description SENSOR DATA
Содержание Proline Prowirl 72
Страница 4: ...QUICK SETUP for quick commissioning Proline Prowirl 72 4 Endress Hauser...
Страница 20: ...3 Installation Proline Prowirl 72 20 Endress Hauser...
Страница 42: ...6 Commissioning Proline Prowirl 72 42 Endress Hauser...
Страница 44: ...7 Maintenance Proline Prowirl 72 44 Endress Hauser...
Страница 114: ...www endress com worldwide BA084D 06 en 12 05 71008404 FM SGML 6 0...