5 Operation
Proline Prowirl 72
34
E Hauser
14
Read sensor information on
the primary process variable
Access type = Read
None
– Byte 0-2: serial number of the sensor
– Byte 3: HART unit ID of the sensor limits and measuring range of
the primary process variable
– Byte 4-7: upper sensor limit
– Byte 8-11: lower sensor limit
– Byte 12-15: minimum span
!
Note!
• The data relate to the primary process variable (= flow).
• Manufacturer-specific units are represented using the HART unit
ID “240”.
15
Read output information of
the primary process variable
Access type = Read
None
– Byte 0: alarm selection ID
– Byte 1: ID for transfer function
– Byte 2: HART unit ID for the set measuring range of the primary
process variable
– Byte 3-6: end of measuring range, value for 20 mA
– Byte 7-10: start of measuring range, value for 4 mA
– Byte 11-14: attenuation constant in [s]
– Byte 15: ID for write protection
– Byte 16: ID for OEM dealer, 17 = E+H
Primary process variable = flow
!
Note!
Manufacturer-specific units are represented using the HART unit
ID “240”.
16
Read the device production
number
Access type = Read
None
Byte 0-2:
production number
17
Write user message
Access = Write
You can save any 32-character long text in the
device with this parameter:
Byte 0-23:
desired user message
Displays the current user message in the device:
Byte 0-23:
current user message in the device
18
Write TAG, TAG description
and date
Access = Write
You can save an 8-character TAG, a 16-
character TAG description and a date with this
parameter:
– Byte 0-5:
TAG
– Byte 6-17:
TAG description
– Byte 18-20:
date
Displays the current information in the device:
– Byte 0-5: TAG
– Byte 6-17: TAG description
– Byte 18-20: date
Command no.
HART command / access type
Command data
(numeric data in decimal form)
Response data
(numeric data in decimal form)
Содержание Proline Prowirl 72
Страница 4: ...QUICK SETUP for quick commissioning Proline Prowirl 72 4 Endress Hauser...
Страница 20: ...3 Installation Proline Prowirl 72 20 Endress Hauser...
Страница 42: ...6 Commissioning Proline Prowirl 72 42 Endress Hauser...
Страница 44: ...7 Maintenance Proline Prowirl 72 44 Endress Hauser...
Страница 114: ...www endress com worldwide BA084D 06 en 12 05 71008404 FM SGML 6 0...