BA 082D/06/en/04.03
No. 50104788

Valid as of software version:
V 1.00.00 (amplifier)

PROline promag 10

Electromagnetic Flow 
Measuring System

Operating Instructions

Содержание Proline Promag 10

Страница 1: ...BA 082D 06 en 04 03 No 50104788 Valid as of software version V 1 00 00 amplifier PROline promag 10 Electromagnetic Flow Measuring System Operating Instructions...

Страница 2: ...e carried out separately Empty pipe full pipe adjustment for Empty Pipe Detection Customer specific configuration Page 49 ff Complex measurement tasks require the configuration of additional functions...

Страница 3: ...ing 35 4 1 Connecting the remote version 35 4 1 1 Connecting Promag W P H 35 4 1 2 Cable specifications 39 4 2 Connecting the measuring unit 40 4 2 1 Transmitter aluminium 40 4 2 2 Terminal assignment...

Страница 4: ...vice Functions 92 11 1 Illustration of the Promag 10 function matrix 92 11 2 Group SYSTEM UNITS 93 11 3 Group OPERATION 94 11 4 Group USER INTERFACE 95 11 5 Group TOTALIZER 96 11 6 Group CURRENT OUTPU...

Страница 5: ...The device must be operated by persons authorised and trained by the facility s owner operator Strict compliance with the instructions in these Operating Instructions is mandatory Endress Hauser will...

Страница 6: ...tic burns etc due to inadequate cleaning will be charged to the owner operator 1 5 Notes on safety conventions and icons The devices are designed to meet state of the art safety requirements They have...

Страница 7: ...he specifications on the order confirmation for the meanings of the individual letters and digits 2 Power supply frequency 11 40 V DC 20 28 V AC Power consumption 6W 8VA 3 Additional information EPD M...

Страница 8: ...316L 6 Additional information examples EPD MSU with Empty Pipe Detection electrode R B with reference electrode EME AWE with exchangeable measuring electrodes 0 5 CAL with 0 5 calibration 7 Permitted...

Страница 9: ...rement Control Regulation and Laboratory Procedures The measuring system described in these Operating Instructions thus complies with the statutory requirements of the EC Directives Endress Hauser con...

Страница 10: ...The wooden covers mounted on the flanges before the device leaves the factory protect the linings on the flanges during storage and transportation Do not remove these covers until immediately before t...

Страница 11: ...to protect it reliably against impact for storage and transportation The original packaging provides optimum protection The permissible storage temperature is 10 50 C preferably 20 C Do not remove th...

Страница 12: ...ing Directly upstream a free pipe outlet in a vertical pipeline Fig 5 Installation location Installation of pumps Do not install the sensor on the intake side of a pump This precaution is to avoid low...

Страница 13: ...to install a cleaning valve Fig 7 Installation in partially filled pipes Down pipes Install a siphon or a vent valve downstream of the sensor in down pipes longer than 5 meters This precaution is to...

Страница 14: ...ation Horizontal orientation The measuring electrode plane should be horizontal This prevents brief insulation of the two measuring electrodes by entrained air bubbles Caution Empty Pipe Detection fun...

Страница 15: ...asuring accuracy Inlet run 5 x DN Outlet run 2 x DN Fig 11 Inlet and outlet run 3 2 4 Vibrations Secure the piping and the sensor if vibration is severe Caution It is advisable to install sensor and t...

Страница 16: ...IN EN 545 double flange reducers can be used to install the sensor in larger diameter pipes The resultant increase in the rate of flow improves measuring accuracy with very slow moving fluids The nomo...

Страница 17: ...scale value Current output v 2 5 m s Pulse value 2 pulse s Low flow cut off v 0 04 m s 25 1 9 300 dm3 min 75 dm3 min 0 50 dm3 1 dm3 min 32 1 1 4 15 500 dm3 min 125 dm3 min 1 00 dm3 2 dm3 min 40 1 1 2...

Страница 18: ...in 2 1 2 65 16 500 gal min 130 gal min 1 gal 2 0 gal min 3 80 24 800 gal min 200 gal min 2 gal 2 5 gal min 4 100 40 1250 gal min 300 gal min 2 gal 4 0 gal min 5 125 60 1950 gal min 450 gal min 5 gal 7...

Страница 19: ...300 m3 h 0 05 m3 5 0 m3 h 250 10 55 1700 m3 h 500 m3 h 0 05 m3 7 5 m3 h 300 12 80 2400 m3 h 750 m3 h 0 10 m3 10 m3 h Flow rate characteristic values Promag P US units Nominal diameter Recommended flow...

Страница 20: ...00 dm3 8 dm3 min 80 3 90 3000 dm3 min 750 dm3 min 5 00 dm3 12 dm3 min 100 4 145 4700 dm3 min 1200 dm3 min 10 00 dm3 20 dm3 min Flow rate characteristic values Promag H US units Nominal diameter Recomm...

Страница 21: ...lectrical machines and switching elements Ensure potential equalisation between sensor and transmitter if necessary The permissible cable length Lmax depends on the fluid conductivity Fig 15 For all f...

Страница 22: ...ctions when installing seals Hard rubber lining additional seals are always necessary Polyurethane lining additional seals are recommended For DIN flanges use only seals according to DIN 2690 Make sur...

Страница 23: ...Hard rubber lining Additional seals must be mounted between the sensor and ground disk as well as between the ground disk and the pipe flange Polyurethane lining Additional seals must be mounted betwe...

Страница 24: ...pply only to pipes not subjected to tensile stress Promag W nominal diameter DIN pressure rating Screws Max torque Nm mm bar Hard rubber Polyurethane 25 PN 40 4 x M 12 15 32 PN 40 4 x M 16 24 40 PN 40...

Страница 25: ...x 7 8 101 75 300 12 Class 150 12 x 7 8 133 103 350 14 Class 150 12 x 1 135 158 400 16 Class 150 16 x 1 128 150 450 18 Class 150 16 x 1 1 8 204 234 500 20 Class 150 20 x 1 1 8 183 217 600 24 Class 150...

Страница 26: ...necessary screw tightening torques onPage 28 ff The mounting of additional ground disks is described on Page 27 Fig 18 Installing the Promag P sensor Seals Comply with the following instructions when...

Страница 27: ...nical Information on Promag 10 separate documents Page 91 Additional seals must be mounted between the ground disk and the pipe flange 1 Place the ground disk and additional seal s between the instrum...

Страница 28: ...DIN pressure rating Screws Max torque Nm mm bar PTFE 25 PN 40 4 x M 12 26 32 PN 40 4 x M 16 41 40 PN 40 4 x M 16 52 50 PN 40 4 x M 16 65 65 PN 16 4 x M 16 87 80 PN 16 8 x M 16 53 100 PN 16 8 x M 16 57...

Страница 29: ...ressure rating Screws Max torque Nm mm PTFE 25 20K 4 x M 16 32 32 20K 4 x M 16 38 40 20K 4 x M 16 41 50 20K 8 x M 16 27 65 10K 4 x M 16 74 80 10K 8 x M 16 38 100 10K 8 x M 16 47 125 10K 8 x M 20 80 15...

Страница 30: ...SO 228 DIN 2999 B DN 2 25 process connections with aseptic moulded seal Weld nipples Tri Clamp L14 AM7 threaded joint DIN 11851 DIN 11864 1 SMS 1145 only DN 25 flange DIN 11864 2 C DN 40 100 process c...

Страница 31: ...could damage the installled seal It is therefore advisable to remove the sensor and the seal prior to welding The pipe has to be spread approximately 8 mm to permit disassembly Cleaning with pigs If p...

Страница 32: ...ch direction and reset it onto the retaining rails 4 Screw the cover of the electronics compartment firmly back onto the transmitter housing Fig 22 Rotating the local display 3 3 6 Affixing the local...

Страница 33: ...sibility Lack of space Extreme fluid ambient temperatures temperature ranges Page 81 Severe vibration 2 g 2 h per day 10 100 Hz Caution If the device is mounted to warm piping make certain that the ho...

Страница 34: ...e of the measuring electrode axis correct Horizontal Is the position of the Empty Pipe Detection EPD electrode correct see Page 14 Were all threaded fasteners tightened to the specified torques when t...

Страница 35: ...tion compartment 2 Sensor remove cover b from the connection housing 3 Feed the signal cable c and coil current cable d through the appropriate cable entries Caution Make sure the connecting cables ar...

Страница 36: ...connected insulated cable shields Note The cable shieldings in the Promag H transmitter are grounded by the strain relief clamps Pay attention to the notes for the cable termination on Page 38 6 5 5...

Страница 37: ...the cable end sleeves do not touch the wire shield on the sensor side Minimum distance 1 mm exception GND green cable Coil current cable Insulate one core of the three core wire at the level of the c...

Страница 38: ...ance 1 mm exception GND green cable Coil current cable Insulate one core of the three core wire at the level of the core reinforcement you only require two cores for the connection On the sensor side...

Страница 39: ...r resistance 50 km Capacitance core shield 420 pF m Operating temperature 20 80 C Cable cross section max 2 5 mm2 Fig 26 Cable cross section a Signal cable b Coil current cable 1 Core 2 Core insulatio...

Страница 40: ...nnection compartment 3 Feed the power supply cable a and signal cable b through the appropriate cable entries 4 Remove terminal connectors d e from the transmitter housing and connect the cables for t...

Страница 41: ...ariant 24 25 26 27 10 A Pulse output Current output HART Pulse output passive Open collector galvanically isolated 30 V DC 250 mA Pulse mode of operation pulse frequency max 100 Hz Status mode of oper...

Страница 42: ...See also the documentation issued by the HART Communication Foundation and in particular HCF LIT 20 HART a technical summary Fig 28 Electrical connection of the HART terminal 1 HART communicator termi...

Страница 43: ...de is standard Promag H No reference electrode present There is always an electrical connection to the fluid via the metallic process connection Caution For installation in metal pipes it is advisable...

Страница 44: ...of the ground terminal provided for the purpose Fig 31 Caution Please also pay special attention to company internal grounding concepts Note The ground cable for flange to flange connections can be or...

Страница 45: ...lectrodes are made of different material Please also pay special attention to company internal grounding concepts Fig 32 Potential equalisation ground disks for plastic or lined pipes Lined pipes cath...

Страница 46: ...ghten the cable entries Fig 34 The cables must loop down before they enter the cable entries water trap Fig 34 This arrangement prevents moisture penetrating the entry Always install the measuring dev...

Страница 47: ...0 Hz Do the cables comply with the specifications see Page 78 Do the cables have adequate strain relief Cables correctly segregated by type Without loops and crossovers Are the power supply and signal...

Страница 48: ...isplay shows measured values dialog texts fault messages and notice messages The display as it appears during standard measuring mode is known as the HOME position operating mode Top line shows main m...

Страница 49: ...e individual functions for operating and configuring the measuring device Example of how to configure a function in this case changing the language for the UI 1 HOME position F Enter the function matr...

Страница 50: ...onfigured values remain safely stored in the EEPROM 5 2 2 Enabling the programming mode The function matrix can be disabled Disabling the function matrix rules out the possibility of inadvertent chang...

Страница 51: ...tion has no effect on the inputs or outputs of the measuring device Fault message Displayed as Lightning flash error type S system error P process error The error in question has a direct effect on th...

Страница 52: ...iles available to the user to provide the following operating aids and programs HART Communicator DXR 275 Selecting device functions with a HART Communicator is a process involving a number of menu le...

Страница 53: ...ables are assigned to the following device variables Primary process variable PV Volume flow Second process variable SV Totalizer Third process variable TV Not assigned Fourth process variable FV Not...

Страница 54: ...e type ID 69 Promag 10 Byte 3 number of preambles Byte 4 universal commands rev no Byte 5 rev no device specific commands Byte 6 software revision Byte 7 hardware revision Byte 8 additional device inf...

Страница 55: ...40 6 Set HART shortform address Access type Write Byte 0 desired address 0 15 Factory setting 0 Note With an address 0 multidrop mode the current output of the primary process variable is set to 4 mA...

Страница 56: ...f measuring range value for 20 mA Bytes 7 10 start of measuring range value for 4 mA Bytes 11 14 attenuation constant in s Byte 15 write protection ID Byte 16 OEM dealer ID 17 E H Factory setting Prim...

Страница 57: ...ng range value for 20 mA Bytes 5 8 start of measuring range value for 4 mA Note Manufacturer specific units are represented using the HART unit ID 240 38 Device status reset configuration changed Acce...

Страница 58: ...the device variable are transferred Byte 0 device variable ID Byte 1 HART unit ID ID of the supported device variables See data on Page 53 Note If the written unit does not suit the device variable th...

Страница 59: ...Error when accessing data of the amplifier EEPROM 3 3 111 Totalizer checksum error 5 0 321 Coil current of sensor is outside tolerance 7 3 351 Current output the current flow is outside the set range...

Страница 60: ...ecks see Page 47 have been successfully completed it is time to switch on the supply voltage The device is ready for operation The measuring device performs a number of internal test functions after p...

Страница 61: ...Select engineering units Volume flow Page 93 Totalizer Page 93 Configure outputs Current output Pulse status output Current span Page 97 Operating mode Page 99 Full scale value Page 98 Pulse value Pa...

Страница 62: ...once a serial number has been entered and stored If a parameter is entered incorrectly during the setup this must be corrected in the relevant function via the function matrix The appropriate informat...

Страница 63: ...ng devices always use cleaning agents that do not attack the surface of the housing and the seals Seals The seals of the Promag H sensor must be replaced periodically particularly in the case of gaske...

Страница 64: ...ound disk for Promag W P Ground disk for potential equalisation DK5GD Mounting kit for remote version aluminium housing Mounting kit for remote version suitable for pipe and wall mounting DK5WM B Moun...

Страница 65: ...Commissioning maintenance analysis Measuring device configuration Service functions Visualisation of process data Trouble shooting Controlling the FieldCheck tester simulator Contact your E H represe...

Страница 66: ...OS buttons simultaneously and switch on the measuring device again The display text will appear in English and is displayed at maximum contrast Measured value indicated but no signal output at the cur...

Страница 67: ...No Cause Remedy spare part S System error Fault message with an effect on the inputs and outputs Notice message without an effect on the inputs and outputs No 0xx Hardware error S CRITICAL FAIL 001 S...

Страница 68: ...frequency is outside the set range 1 Increase the setting for pulse weighting 2 When selecting the pulse width choose a value that can still be processed by a connected counter e g mechanical totalize...

Страница 69: ...OL FLOW 692 Simulation of volume flow active Switch off simulation Typ e Error message No Cause Remedy P Process error Fault message with an effect on the inputs and outputs Notice message without an...

Страница 70: ...1 Perform empty or full pipe adjustment and then switch on Empty Pipe Detection Page 107 2 Remote version check the terminal connections of the EPD cable Page 35 ff 3 Fill the measuring tube The curr...

Страница 71: ...formation on Page 51 ff Current output MIN VALUE 4 20 mA 25 mA 2 mA 4 20 mA NAMUR 3 5 mA 4 20 mA US 3 75 mA 4 20 mA 25 mA HART 2 mA 4 20 mA HART NAMUR 3 5 mA 4 20 mA HART US 3 75 mA MAX VALUE 4 20 mA...

Страница 72: ...rts The illustration below shows the available scope of spare parts Note You can order spare parts directly from your E H service organisation by providing the serial number printed on the nameplates...

Страница 73: ...onics compartment from the transmitter housing 3 Remove the display a from the connection compartment 4 Press the side latches b and flip down the cover of the connection compartment 5 Disconnect the...

Страница 74: ...ectrode signal cable d Connector of the coil current cable e Connector for the power supply f Connector for the current output and pulse status output g Connector for the local display h Threaded fast...

Страница 75: ...rew cover of the electronics compartment from the transmitter housing 3 Press the side latches and flip down the cover of the connection compartment 4 Disconnect the connector a for the power supply 5...

Страница 76: ...re history Software version date Software modification Documentation Modifications Supplements Amplifier V 1 00 00 01 2003 Original software Compatible with FieldTool Commuwin II version 2 05 03 and h...

Страница 77: ...hanical unit Remote version Sensor is mounted separate from the transmitter Transmitter Promag 10 Sensor Promag W DN 25 600 1 2 24 Polyurethane liner for water and waste water applications application...

Страница 78: ...n alarm Current output Failsafe mode can be selected Pulse output Failsafe mode can be selected Status output Not conductive in the event of fault or power supply failure Details Page 71 Load See Outp...

Страница 79: ...istics Reference operating conditions To DIN 19200 and VDI VDE 2641 Fluid temperature 28 C 2 K Ambient temperature 22 C 2 K Warm up time 30 minutes Installation Inlet run 10 x DN Outlet run 5 x DN Sen...

Страница 80: ...device at a shady location Avoid direct sunlight particularly in warm climatic regions The transmitter must be mounted separate from the sensor if both the ambient and fluid temperatures are high Flui...

Страница 81: ...TF Medium temperature Fig 44 Remote versions Promag P TA Ambient temperature TF Medium temperature Promag H Sensor DN 2 25 20 150 C DN 40 100 20 150 C Seal EPDM 20 130 C Viton 20 150 C Kalrez 20 150...

Страница 82: ...olute pressure mbar with different fluid temperatures mm inch 25 C 60 C 80 C 100 C 130 C 150 C 180 C 25 600 1 24 Polyurethane 0 0 65 600 1 24 Hard rubber 0 0 0 Promag P nominal diameter Measuring tube...

Страница 83: ...SI Sensor Transmitter 25 1 PN 40 5 7 Class 150 5 7 PN 40 5 3 Class 150 5 3 3 1 32 1 1 4 6 4 6 0 3 1 40 1 1 2 7 8 7 8 7 4 7 4 3 1 50 2 9 0 9 0 8 6 8 6 3 1 65 2 1 2 PN 16 10 4 PN 16 10 0 3 1 80 3 12 4 1...

Страница 84: ...23 5 23 5 3 1 200 8 PN 10 43 4 43 4 PN 10 43 43 3 1 250 10 63 4 73 4 63 73 3 1 300 12 68 4 108 4 68 108 3 1 Promag transmitter compact version 1 8 kg Weight data valid for standard pressure ratings an...

Страница 85: ...A 105 DN 350 with Al Zn protective coating DN 300 with Amerlock 400 coating JIS S20C DN 350 with Al Zn protective coating DN 300 with Amerlock 400 coating Ground disks 1 4435 316L or Alloy C 22 Elect...

Страница 86: ...ation Promag 10 H TI 095D 06 en Fitted electrodes Promag W Measuring electrodes reference electrodes and Empty Pipe Detection electrodes Available as standard with 1 4435 316L Alloy C 22 Promag P Meas...

Страница 87: ...23 EC Pressure Equipment Directive For larger nominal diameters certified flowmeters to Category III are optionally also available if necessary depends on fluid and operating pressure All devices are...

Страница 88: ...us accessories which can be ordered separately from Endress Hauser are available for the transmitter and the sensor see Page 64 Your E H service organisation can provide detailed information on the or...

Страница 89: ...20K 24 32 PN 40 20K 32 40 1 1 2 PN 40 Cl 150 20K 38 50 2 PN 40 Cl 150 20K 50 65 PN 16 10K 66 66 80 3 PN 16 Cl 150 10K 76 79 100 4 PN 16 Cl 150 10K 102 102 125 PN 16 10K 127 127 150 6 PN 16 Cl 150 10K...

Страница 90: ...K 202 250 10 PN 10 Cl 150 10K 256 300 12 PN 10 Cl 150 10K 306 Promag H nominal diameter Pressure ratings Measuring tube internal diameter DIN mm ANSI inch DIN bar PFA 2 1 12 PN 16 PN 40 2 25 4 5 32 PN...

Страница 91: ...nsions can be found in the following documents Technical Information Promag 10 W TI 093D 06 en Technical Information Promag 10 P TI 094D 06 en Technical Information Promag 10 H TI 095D 06 en You can a...

Страница 92: ...ESS CODE P 94 CONTRAST LCD P 95 OVERFLOW P 96 VALUE 20 mA P 98 PULSE VALUE P 99 ASSIGN STATUS P 101 TAG DESCRIPTION P 105 EPD MODE P 107 MEASURING MODE P 108 ZERO POINT P 111 ALARM DELAY P 113 SIM MEA...

Страница 93: ...Mgal min Mgal h Mgal day Barrel normal fluids 31 5 gal bbl bbl s bbl min bbl h bbl day Barrel beer 31 0 gal bbl bbl s bbl min bbl h bbl day Barrel petrochemicals 42 0 gal bbl bbl s bbl min bbl h bbl...

Страница 94: ...hen programming is disabled You can enable programming by entering your private code factory setting 10 see DEF PRIVATE CODE function on Page 94 User input Max 4 digit number 0 9999 Note The programmi...

Страница 95: ...ue and the engineering unit e g 1 2 l h indicating that the measuring system is computing with more decimal places than can be shown on the display CONTRAST LCD Use this function to optimise display c...

Страница 96: ...in the UNIT VOLUME function Page 93 OVERFLOW The total for the totalizer s overflow aggregated since measuring commenced appears on the display Total flow quantity is represented by a floating point...

Страница 97: ...HART NAMUR 4 20 mA US 4 20 mA HART US Factory setting 4 20 mA 25 mA HART Current span work range and signal on alarm level A Current span option Work range Lower signal on alarm level Upper signal on...

Страница 98: ...n Factory setting Depends on nominal diameter and country value dm3 m3 or US gal US Mgal Corresponds to the factory setting for the full scale value see Page 116 ff Note The appropriate unit is taken...

Страница 99: ...ERATION MODE function Use this function to define the flow at which a pulse is triggered These pulses can be totalled by an external totalizer and the total flow quantity since measuring started can b...

Страница 100: ...n external totalizer e g mechanical totalizer PLC etc The pulse output s response to faults is defined in the central FAILSAFE MODE function Page 112 Caution Buffering pulse memory takes place if the...

Страница 101: ...an the switch off point Positive and negative values are permitted User input 5 digit floating point number unit Factory setting 0 unit Note The appropriate unit is taken from the SYSTEM UNITS group s...

Страница 102: ...int entered is 1 m3 h the status output switches off at 1 m3 h not conductive and switches on again at 1 m3 h conductive Set the switch point to 0 if your process calls for direct switchover no switch...

Страница 103: ...tem OK con ductive System or process error Fault Response to error outputs Inputs and totalizers not con ductive Notice message System OK con ductive System or process error Notice Continuation of mea...

Страница 104: ...104 Endress Hauser Flow direction Forward con ductive Reverse not con ductive Limit value volume flow Limit value not overshot or undershot con ductive Limit value overshot or undershot not con ductiv...

Страница 105: ...tag description at the local display or via the HART protocol User input Max 16 character text permitted characters are A Z 0 9 punctuation marks Factory setting _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ no tex...

Страница 106: ...5 digit floating point number unit Factory setting Depends on nominal diameter and country value dm3 m3 or US gal Corresponds to the factory setting for the low flow cut off see Page 116 ff Note The...

Страница 107: ...es on Empty Pipe Detection EPD Flow cannot be measured correctly unless the measuring pipe is completely full This status can be monitored at all times by means of the EPD An empty or partially filled...

Страница 108: ...ponses of the individual outputs and the internal totalizer in each of the measuring modes are described in detail on the following pages Current output STANDARD Only the flow components for the selec...

Страница 109: ...signal switches at the defined switch points irrespective of the sign In other words if you define a switch point with a positive sign the status output signal switches as soon as the value is reache...

Страница 110: ...ptions OFF ON Signal output is set to the ZERO FLOW value Factory setting OFF SYSTEM DAMPING Use this function to set the filter depth of the digital filter This reduces the sensitivity of the measuri...

Страница 111: ...d calibration Note This value is also provided on the sensor nameplate ZERO POINT This function shows the current zero point correction value for the sensor Zero point correction is determined and set...

Страница 112: ...NIMUM VALUE The current output is set to the lower value of the signal on alarm level The values in question can be found in the CURRENT SPAN function on Page 97 MAXIMUM VALUE The current output is se...

Страница 113: ...e delayed by the time corresponding to the setting before being forwarded to the higher order controller process controller etc It is therefore imperative to check in advance in order to make sure whe...

Страница 114: ...her they respond correctly During this time the words SIMULATION MEASURAND appear on the display Options OFF VOLUME FLOW Factory setting OFF Caution The measuring device cannot be used for measuring w...

Страница 115: ...ERSION Function description SENSOR VERSION SERIAL NUMBER Use this function to view the serial number of the sensor SENSOR TYPE Use this function to view the sensor type Function description AMPLIFIER...

Страница 116: ...300 dm3 min 2 50 dm3 dm3 65 2 1 2 8 dm3 min 500 dm3 min 5 00 dm3 dm3 80 3 12 dm3 min 750 dm3 min 5 00 dm3 dm3 100 4 20 dm3 min 1200 dm3 min 10 00 dm3 dm3 125 5 30 dm3 min 1850 dm3 min 15 00 dm3 dm3 1...

Страница 117: ...25 0 25 gal min 18 gal min 0 20 gal gal 1 1 4 32 0 50 gal min 30 gal min 0 20 gal gal 1 1 2 40 0 75 gal min 50 gal min 0 50 gal gal 2 50 1 25 gal min 75 gal min 0 50 gal gal 2 1 2 65 2 0 gal min 130 g...

Страница 118: ...11 Description of Device Functions PROline Promag 10 118 Endress Hauser...

Страница 119: ...Control Function matrix 49 Current output Electrical connection 41 Technical data 78 Current span 97 D Declaration of conformity CE mark 9 Declaration of contamination for return of measuring devices...

Страница 120: ...e to errors 71 Installation See Installation instructions Installation conditions Installation conditions Adapters 16 Dimensions 12 Down pipes 13 Inlet and outlet run 15 Installation location 12 Insta...

Страница 121: ...te code 94 Process connections 86 Process error 51 Process error messages 69 Process errors without display messages 70 Programming mode Disable 50 Enable 50 Pulse value 99 Pulse width 100 Pulse statu...

Страница 122: ...smitter Electrical connection 40 Length of connecting cable remote version 21 Rotating the transmitter housing 31 Trouble shooting 66 Types of error system and process errors 51 U Unit Volume 93 Volum...

Страница 123: ...f health o biological hazardous o inflammable o safe Please mark the appropriate warning hints Reason for return ___________________________________________________________________________ ___________...

Страница 124: ...Costa Rica San Jose Euro Tec Costa Rica S A Tel 2 20 28 08 Fax 2 96 15 42 Ecuador Quito Insetec Cia Ltda Tel 02 2 26 91 48 Fax 02 2 46 18 33 El Salvador San Salvador Automatizacion y Control Industria...