Tankside Monitor NRF81
Diagnostics and troubleshooting
Hauser
95
Diagnostic
number
Short text
Remedy instructions
Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
584
V1 C1-4 user value source
invalid
Set valid user value input
selector
C
Warning
585
Simulation distance
Deactivate simulation
C
Warning
585
V1 D1-4 user value source
invalid
Set valid user value input
selector
C
Warning
586
Record map
Recording of mapping
please wait
C
Warning
586
V1 A1-4 discrete value
source invalid
Set valid user discrete input
selector
C
Warning
587
V1 B1-4 discrete value
source invalid
Set valid user discrete input
selector
C
Warning
588
V1 C1-4 discrete value source
invalid
Set valid user discrete input
selector
C
Warning
589
V1 D1-4 discrete value
source invalid
Set valid user discrete input
selector
C
Warning
590
Modbus A1-4 percent source
invalid
Set valid percentage input
selector
C
Warning
591
Modbus B1-4 percent source
invalid
Set valid percentage input
selector
C
Warning
592
Modbus C1-4 percent source
invalid
Set valid percentage input
selector
C
Warning
593
Modbus D1-4 percent source
invalid
Set valid percentage input
selector
C
Warning
594
V1 A1-4 percent source
invalid
Set valid percentage input
selector
C
Warning
595
V1 B1-4 percent source
invalid
Set valid percentage input
selector
C
Warning
596
V1 C1-4 percent source
invalid
Set valid percentage input
selector
C
Warning
597
V1 D1-4 percent source
invalid
Set valid percentage input
selector
C
Warning
598
DIO A1-2 source no longer
valid
Change input source
C
Warning
599
DIO A3-4 source no longer
valid
Change input source
C
Warning
Diagnostic of process
801
Energy too low
Increase supply voltage
S
Warning
803
Current loop
1. Check device configuration.
2. Check wiring.
F
Alarm
803
Current loop 1 to 2
M
Warning
803
Current loop
C
Warning
825
System temperature
1. Check ambient temperature
2. Check process temperature
S
Warning
825
System temperature
F
Alarm
826
Sensor temperature
1. Check ambient temperature
2. Check process temperature
S
Warning
826
Sensor temperature
F
Alarm
844
Process value out of
specification
1. Check process value
2. Check application
3. Check sensor
S
Warning
844
Process value out of
specification
S
Warning