Operating menu
Tankside Monitor NRF81
158
Hauser
User entry
0 to 255
Factory setting
1
Additional information
Read access
Operator
Write access
Maintenance
Level mapping
Navigation
Setup → Advanced setup → Communication → V1 X1-4 → Configuration → Level
mapping
Prerequisite
Communication interface protocol (→ 153)
V1
Description
Determines the transmittable range of levels.
Selection
• +ve
• +ve & -ve
Factory setting
+ve
Additional information
Read access
Operator
Write access
Maintenance
In V1, the level is always represented by a number in the range from 0 to 999
999. This
number corresponds to a level as follows:
"Level mapping" = "+ve"
Number
Corresponding level
0
0.0 mm
999
999
99
999.9 mm
"Level mapping" = "+ve & -ve"
Number
Corresponding level
0
0.0 mm
500
000
50
000.0 mm
500
001
–0.1 mm
999
999
–49
999.9 mm
Содержание NRF81
Страница 250: ...www addresses endress com 71406899 71406899...