Commissioning
Micropilot S FMR531 with HART/4...20 mA
50
E Hauser
Display "dist./meas.value" (008)
The
distance
measured from the reference point to the product surface and the
level
calculated
with the aid of the empty alignment are displayed again. Check whether the values correspond to
the actual level or the actual distance. The following cases can occur:
• Distance correct – level correct
→
continue with the next function, "
check distance
"
(051)
.
• Distance correct . level incorrect
→
Check "
empty calibr
"
(005)
• Distance correct – level incorrect
→
continue with the next function, "
check distance
"
(051)
.
Function "history reset" (009)
By this function a history reset of the device is performed, i.e. the correspondance table between
level an index values is deleted. A new correspondance table will be filled and stored after the
history reset, cf. Page 51.
"
Caution!
A history reset must be performed after:
• first installation or
• change of basic setup or
• change of the installation situation.
In this case also effect a reset of the dip table in function "
dip table mode
"
(033)
.
⇒