Trouble-shooting
Micropilot S FMR531 with HART/4...20 mA
80
E Hauser
Modification nameplate
When ordering parts that are listed in the product structure (see Page 8) , a check must be made as
to whether the instrument description on the nameplate is still valid, e.g. for:
• an antenna component,
• an electronics module
If the instrument designation on the nameplate has changed then a modified nameplate must also
be ordered. The information about the new instrument must then be entered on the modified
nameplate. This must then be attached to the housing of the Micropilot S. See instructions packed
with nameplate.
Housing T12
The complete order code must be specified when ordering the replacement housing so that the
correct nameplate can be delivered, e.g.
• FMR531-A4VCW2AA2A
You must label the nameplate yourself.
"
Caution!
• It is not possible to convert a standard instrument into an Ex instrument by simply exchanging
parts.
• When repairing certified instruments, the relevant regualtions must be followed.
• For FM approved instruments, it is forbidden to make any changes to the instrument that are not
expressly authorised in the operating manual. Contravening this prohibition can invalidate the
approval for operation of the instrument.