Micropilot S FMR531 with HART/4...20 mA
Commissioning
E Hauser
53
Display "custody mode" (0A9)
This indicates the instrument calibration mode. The calibration mode (active) can be set using the
hardware security lock on the electronics (see Page 32).
Selection:
•
inactive
• active pos.
• active neg.
active pos.
The custody mode (instrument is lead-sealed and accurate to the nearest mm) is active and is held.
active neg.
Custody mode (instrument is lead-sealed and accurate to the nearest mm) is activated and not held,
e.g. because the signal-to-noise ratio is less than 5 dB (refer to
function "
echo quality
"
(056)
in the function group "
extended calibr.
"
(05)
).
"
Caution!
After entering all the values and completing mounting and aligning work, enter the Reset Code
"
555
" in the function "
reset
"
(0A3)
to reset the instrument history for auto-correction or set history
reset to "
yes
" in order to reset the instrument history for the auto-correction.
⇒