Mounting
Micropilot S FMR531 with HART/4...20 mA
18
E Hauser
Measuring range
The usable measuring range depends on the size of the antenna, the reflectivity of the medium, the
mounting location, and eventual interference reflections.
The following tables describe the groups of media as well as the achievable measuring range as a
function of application and media group. If the dielectric constant of a medium is unknown, it is
recommended to assume media group B to ensure a reliable measurement.
Measuring range depending on product class for Micropilot S FMR531
Media group
DC (
ε
r)
Examples
A
1.4...1.9
non-conducting liquids, e.g. liquefied gas (LPG). For more information please contact
your Hauser representative
.
B
1.9...4
non-conducting liquids, e.g. benzene, oil, toluene, white products, black products,
crudes, bitumen/asphalts, …
C
4...10
e.g. concentrated acids, organic solvents, esters, aniline, alcohol, acetone, …
D
> 10
conducting liquids, e.g. aqueous solutions, dilute acids and alkalis
Media group
Free space (Storage tank)
Measuring range
DN150/200/250
A
DC (
ε
r) = 1.4...1.9
—
B
DC (
ε
r) = 1.9...4
10 m/33 ft
C
DC (
ε
r) = 4...10
DN150: 15 m/50 ft
D
DC (
ε
r) > 10
DN150: 20 m/65 ft
max. measuring
range with custody
transfer approvals
NMi and PTB:
10 m/33 ft
Содержание Micropilot S FMR531
Страница 4: ...Brief overview Micropilot S FMR531 with HART 4 20 mA 4 Endress Hauser ...
Страница 87: ...Micropilot S FMR531 with HART 4 20 mA Technical data Endress Hauser 87 ...
Страница 98: ...Micropilot S FMR531 with HART 4 20 mA Index 98 Endress Hauser ...
Страница 100: ...www endress com worldwide BA207F 00 en 04 09 71093754 CCS FM SGML 6 0 ProMoDo 71093754 ...