Micropilot S FMR531 with HART/4...20 mA
Commissioning
E Hauser
47
Function "full calibr." (006)
This function is used to enter the distance from the minimum level to the maximum level (=span).
L00-FMR2xxxx-14-00-06-en-009
In principle, it is possible to measure up to the tip of the antenna. However, due to considerations
regarding corrosion and build-up, the end of the measuring range should not be chosen any closer
than 50 mm (2”) to the tip of the antenna.
!
Note!
If
bypass
or
stilling well
was selected in the "
tank shape
"
(002)
function, the pipe diameter is
requested in the following step.
Function "pipe diameter" (007)
This function is used to enter the pipe diameter of the stilling well or bypass pipe.
L00-FMR2xxxx-14-00-00-de-011
⇒
E HAUSER
E
+
–
⇒
E HAUSER
E
+
–
100%
0%
ø
100%
0%
ø