Commissioning
Micropilot S FMR531 with HART/4...20 mA
58
E Hauser
store point
The displayed value pair is sorted in the table.
!
Note!
For sorting, the following criteria must be met:
• Measured variables may not be equal but have different dip values.
• A measured variable available in the table is recognised as equal when it is closer than 1 mm to
the sorting value.
• After successful sorting, the setting remains at "
edit point
" and the number of free table entries
is decremented.
"
Caution!
If the value cannot be sorted, the setting remains at the previous menu option. No
warning or error message is generated. However, the number of remaining table entries is not
decremented.
delete point
The currently displayed point is deleted from the table. After deletion, the previous point is
displayed. If the table only consisted of one point before deletion, then the current measured
variable is displayed as a value pair.
return
By selecting this point, you return to the function "
dip table mode
"
(033)
.
next point
This scrolls down in the table. If the table is empty, you can still select this option. However, the
displayed value does not change.
previous point
This scrolls up in the table. If the table is empty, you can still select this option. However, the
displayed value does not change.
"
Caution!
After entering one or more points into the dip table, make sure that the dip table is activated in the
"
table on
" dip table mode.
Содержание Micropilot S FMR531
Страница 4: ...Brief overview Micropilot S FMR531 with HART 4 20 mA 4 Endress Hauser ...
Страница 87: ...Micropilot S FMR531 with HART 4 20 mA Technical data Endress Hauser 87 ...
Страница 98: ...Micropilot S FMR531 with HART 4 20 mA Index 98 Endress Hauser ...
Страница 100: ...www endress com worldwide BA207F 00 en 04 09 71093754 CCS FM SGML 6 0 ProMoDo 71093754 ...