Memocheck Sim CYP03D
Operation
Hauser
41
8.2.14 Extended functions: Nitrate
Path: Menu/Advanc. function
Function
Configuration options
(factory settings in bold)
Calibration value
Temp. offset
-10.0 to 10.0 ˚C
0.0 ˚C
Error simulation
Flash lamp defect
Turbidit. too high
Electronics test
Filter change
Meas. val. uncert.
No cal. data
Temperature err.
Measurement stop
Meas. value inval.
Once you have selected an error using ,
the error is output at the transmitter. To
delete the error again, remove the check
mark again using .
Ramp
Ramp
(continuous)
Ramp (gradual)
Sim. main value
Starting
value
Stop value
Duration
Number of
steps
Time per step
Content NO3
0.001 mg/l to
550 mg/l
0.001 mg/l
0.001 mg/l to
550 mg/l
550 mg/l
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Content NO3-N
0.0 µg/l to
500.0 mg/l
0.0 µg/l
0.0 µg/l to
500.0 mg/l
500.0 mg/l
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Temperature
-40.0 to
150.0 ˚C
-40.0 °C
-40.0 to
150.0 ˚C
150.0 °C
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Raw value temp.
-40.0 to
150.0 ˚C
-40.0 °C
-40.0 to
150.0 ˚C
150.0 °C
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s