Memocheck Sim CYP03D
Operation
Hauser
39
8.2.12 Extended functions: Chlorine
Path: Menu/Advanc. function
Function
Configuration options
(factory settings in bold)
Calibration value
Slope
-500.0 to -0.01 nA/g/l
-25.0 nA/g/l
Zero point
-100.0 nA to 100.0 nA
0.0 pA
Temp. offset
-10.0 to 10.0 ˚C
0.0 ˚C
Temp. gradient
-3.0 to 3.0
1.0
Error simulation
Temp. sens. defect
Sensor supply
Once you have selected an error using ,
the error is output at the transmitter. To
delete the error again, remove the check
mark again using .
Ramp
Ramp
(continuous)
Ramp (gradual)
Sim. main value
Starting
value
Stop value
Duration
Number of
steps
Time per step
Chlor. concentr.
0.0 to 200.0
mg/l
0.0 mg/l
-0.02 to
120.0 mg/l
120.0 mg/l
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Current
-1.5 µA to 0.0
pA
-1.5 µA
-1.5 µA to 0.0
pA
0.0 pA
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Temperature
-40.0 to 60.0
˚C
-40.0 °C
-40.0 to 60.0
˚C
60.0 °C
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
pH value
-2.0 to 16.0
pH
-2.0 pH
-2.0 to 16.0
pH
16.0 pH
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Raw value current
-1.5 µA to 0.0
pA
-1.5 µA
-1.5 µA to 0.0
pA
0.0 pA
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Raw value temp.
-40.0 to 60.0
˚C
-40.0 °C
-40.0 to 60.0
˚C
60.0 °C
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s