Memocheck Sim CYP03D
Operation
Hauser
21
8.1.3
Selecting parameters
In the setup, you configure which parameters are to be simulated.
1. Path:
Menu
/
Setup
/
Simulation
/
Parameter
.
The parameters currently configured are displayed.
2. Select new parameter:
Parameter
/
Memosens
or
Fixed cable
. Select the desired
parameter.
If the parameter is changed, the following prompt appears:
ESC = cancel
= direct parameter change, the last settings for the current parameter are lost
= save setup, followed by a prompt to specify where to save the setup. Select a
free location so that setups already saved are not overwritten.
3. Select desired option.
Under
Menu
/
Setup
/
Save setup
, you can save up to 10 setups under the relevant
parameter name. If you want to use a saved setup, select it under the "Setup/Load setup"
menu.
8.1.4
Configuring sim. main value and test values
A setup comprises the selected parameter, a simulation main value and test values.
You can change the simulation main value in the simulation mode using a user-definable
delta value. The delta value is the increment by which you change the simulation value by
pressing and .
Test values are all other measured values (including raw measured values) of a test setup. Test
values are set to a fixed value, which cannot be modified in the simulation mode.
1.
Menu
/
Setup
/
Simulation
/
Parameter
/
Memosens
or .../
Fixed cable
: Select the
parameter, e. g.
pH glass
.
2.
Menu
/
Setup
/
Simulation
/
Sim. main value
: Select the measured value to be simulated,
e. g.
pH value
.
A prompt asking if you wish to change the delta value () or not ().
3. Select .
The current delta value is now displayed, e. g. 00.10 pH.
4. Change the current value, e. g. to 00.50 pH.
5. Accept value (use to select the beside the value and then ).