Operation
Memocheck Sim CYP03D
22
Hauser
6.
Menu
/
Setup
/
Simulation
/
Test values
: set other measured values, which are shown on
the display (
Temperature
only) or on the transmitter as a fixed value.
7. Change the value to the desired display, e. g. 25.00 °C.
The test value remains at the value set here and cannot be modified in the
simulation mode.
8. Pressing and for longer takes you directly to the simulation mode.
The
Sim. main value
is displayed as the first value in the simulation mode. You can press
or to change the set delta value. The temperature is displayed as the second value. Pressing
and does not affect this value. The other test values, which are not visible on the
Memocheck display, can only be read off the transmitter or output to transmitter outputs.
Parameter Chlorine
To ensure that the simulator and transmitter display the identical chlorine
concentration, the pH value which is used for calculation purposes must be the same for
the transmitter and the simulator.
Parameter Oxygen (amp.) or Oxygen (opt.Memo.)
To ensure that the simulator and transmitter display the identical oxygen concentration,
the following values used for calculation purposes must be the same for the transmitter
and the simulator:
Salinity
and
Process pressure
/
Altitude
.