Operation
Memocheck Sim CYP03D
26
Hauser
Path: Menu/Advanc. function
Function
Configuration options
(factory settings in bold)
Error simulation
Glass SCS failure
Temp. sens. defect
Glass SCS warning
Once you have selected an error using ,
the error is output at the transmitter. To
delete the error again, remove the check
mark again using .
Ramp
Ramp
(continuous)
Ramp (gradual)
Sim. main value
Starting
value
Stop value
Duration
Number of
steps
Time per step
pH value
-2.0 to 16.0
pH
-2.0 pH
-2.0 to 16.0
pH
16.0 pH
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Temperature
-40.0 to
150.0 ˚C
-40.0 °C
-40.0 to
150.0 ˚C
150.0 °C
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Raw value
-750.0 to
750.0 mV
-750.0 mV
-750.0 to
750.0 mV
750.0 mV
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Raw value temp.
-40.0 to
150.0 ˚C
-40.0 °C
-40.0 to
150.0 ˚C
150.0 °C
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
SCS resistance
0.001 MΩ to
1.0 TΩ
0.001 MΩ
0.001 MΩ to
1.0 TΩ
1.0 TΩ
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
8.2.5
Extended functions: pH Isfet
Path: Menu/Advanc. function
Function
Configuration options
(factory settings in bold)
Calibration value
Temp. offset
-10.0 to 10.0 ˚C
0.0 ˚C
Slope
0.01 to 65.0 mV/pH
59.16 mV/pH
pH comp. isoth.
0.0 to 12.0 pH
7.0 pH
mV comp. isoth.
-300.0 to 300.0 mV
0.0 mV
Asymmetry
-300.0 to 300.0 mV
0.0 mV