Operation
Memocheck Sim CYP03D
40
Hauser
8.2.13 Extended functions: Turbidity
Path: Menu/Advanc. function
Function
Configuration options
(factory settings in bold)
Calibration value
Temp. offset
-10.0 to 10.0 ˚C
0.0 ˚C
Error simulation
LED error
Turbidit. too high
Electronics test
Sensor polluted
Meas. val. uncert.
No cal. data
Temperature err.
Meas. value inval.
Measurement stop
Once you have selected an error using ,
the error is output at the transmitter. To
delete the error again, remove the check
mark again using .
Ramp
Ramp
(continuous)
Ramp (gradual)
Sim. main value
Starting
value
Stop value
Duration
Number of
steps
Time per step
Temperature
-40.0 to
150.0 ˚C
-40.0 °C
-40.0 to
150.0 ˚C
150.0 °C
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
TU value [FNU]
0.0 to
10000.0 FNU
0.0 FNU
0.0 to
10000.0 FNU
10000.0 FNU
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
TU value [g/l]
0.001 to 1000
g/l
0.001 g/l
0.001 to 1000
g/l
1000 g/l
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Raw value temp.
-40.0 to
150.0 ˚C
-40.0 °C
-40.0 to
150.0 ˚C
150.0 °C
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s