Memocheck Sim CYP03D
Operation
Hauser
27
Path: Menu/Advanc. function
Function
Configuration options
(factory settings in bold)
Error simulation
Leak. curr. alarm
Leak. curr. warn
Temp. sens. defect
Sensor supply
Once you have selected an error using ,
the error is output at the transmitter. To
delete the error again, remove the check
mark again using .
Ramp
Ramp
(continuous)
Ramp (gradual)
Sim. main value
Starting
value
Stop value
Duration
Number of
steps
Time per step
pH value
-2.0 to 16.0
pH
-2.0 pH
-2.0 to 16.0
pH
16.0 pH
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Temperature
-40.0 to
150.0 ˚C
-40.0 °C
-40.0 to
150.0 ˚C
150.0 °C
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Raw value
-750.0 to
750.0 mV
-750.0 mV
-750.0 to
750.0 mV
750.0 mV
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Raw value temp.
-40.0 to
150.0 ˚C
-40.0 °C
-40.0 to
150.0 ˚C
150.0 °C
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s